PowerXpert™ Help : Verifying MA243x0A Performance : Required Equipment
 
Required Equipment
Required Equipment for MA243x0A
Equipment Description
Manufacturer and Model
Critical Specifications
Vector Network Analyzer
Anritsu
MS4644B (40GHz)
MS4645B (50GHz)
Reflection Coefficient
UNC 0.015, 10MHz to 50MHz
UNC 0.012, 50MHz to 18GHz
UNC 0.016, 18 to 33GHz
UNC 0.021, 33 to 40GHz
UNC 0.024, 40 to 50GHz
Synthesizer
Anritsu
MG3694C (40GHz)
MG3695C (50GHz)
Output Power ≥ +15 dBm, 50 MHz to 40 GHz
Output Power Setting Resolution: 0.01dBm
Harmonics –40 dBc
Source VSWR 2.0
Reference Power Meter
Anritsu
ML2438A
Instrumentation Accuracy  0.5 %
Reference Power Sensor
Anritsu
MA24004A (40GHz)
MA24005A (50GHz)
NIST Calibration or equivalent
10 dB Attenuator
Anritsu
41KC-10 (40GHz)
41V-10 (50GHz)
VSWR  1.10, 10 MHz to 12 GHz
VSWR  1.15, > 12 GHz to 18 GHz
VSWR  1.18, > 18 GHz to 26.5 GHz
VSWR  1.28, > 26.5 GHz to 40 GHz
Power Splitter
Anritsu
K241C (40GHz
V241C (50GHz)
Effective Output VSWR < 2.0, DC to 18 GHz
Effective Output VSWR < 2.3, 18 GHz to 40 GHz
Effective Output VSWR < 2.6, 40 GHz to 50 GHz
Personal Computer
Any
Type V Calibration Kit
Anritsu
3654D