PowerXpert™ Help : Verifying MA24105A Performance : Required Equipment
 
Required Equipment
Required Equipment
Equipment Description
Manufacturer and Model
Critical Specifications
Vector Network Analyzer (Pretest)
Anritsu MS4642B
or equivalent
Reflection Coefficient
Uncertainty  0.013, 350 MHz to 2 GHz
Uncertainty  0.020, 2 GHz to 4 GHz
Synthesizer
(Cal. Factor and Linearity Tests)
Anritsu MG3692B
or equivalent
Output Power:  +20 dBm 0.05 GHz to 4 GHz
Output Power Setting Resolution: 0.01 dBm
Harmonics:  –40 dBc
Source VSWR  2.00
Reference Power Meter
(Cal. Factor and Linearity Tests)
Anritsu ML2438A
or equivalent
Instrumentation Accuracy  0.5 %
Reference Power Sensor
(Cal. Factor and Linearity Tests)
Anritsu MA24002A
or equivalent
NIST Calibration or equivalent
10 dB N Attenuator
(Linearity Test)
Aeroflex Model 1433
VSWR  1.15, 350 MHz to 4 GHz, 250 W min.
10 dB K Attenuator
(Frequency Response and Linearity Tests)
Anritsu 41KC-10
VSWR  1.15, 350MHz to 4 GHz, 2 W min.
Low Power 30 dB N Attenuator (Linearity Tests)
42N50A-30
VSWR 1.2, 350 MHz to 4 GHz, 30 W min.
50 ohm Termination
Aeroflex Model 1433
VSWR  1.15, 350 MHz to 4 GHz, 250 W min.
Adapter N(f) to K(f)
(Frequency Response and Linearity Tests)
Anritsu 34ANF50, 34AS50, 34AN50, and 34ASF50
VSWR  1.05, 350 MHz to 4 GHz
Power Coupler (Linearity Test)
MITEQ CD2-522-30N
350 MHz to 4.0 GHz, Coupling = 30 dB, 250 W min.
Amplifier (Directivity, Frequency Response and Linearity Tests)
Ophir 5163, 5125
350 MHz to 4.0 GHz, 250 W min.
Harmonics/Spurious < –20 dBc
Personal Computer
Any