PowerXpert™ Help : Verifying MA24106A Performance : Required Equipment
 
Required Equipment
Required Equipment
Equipment Description
Manufacturer and Model
Critical Specifications
Vector Network Analyzer (Pretest)
Anritsu MS4642B
or equivalent
Reflection Coefficient
Uncertainty 0.013, 10 MHz to 2 GHz
Uncertainty 0.020, 2 GHz to 6 GHz
Synthesizer
(Cal. Factor and Linearity Tests)
Anritsu MG3692C
or equivalent
Output Power: >+18 dBm, 50 MHz to 6 GHz
Output Power Setting Resolution: 0.01 dBm
Harmonics: –40 dBc
Source VSWR ≤ 2.00
Reference Power Meter
(Cal. Factor and Linearity Tests)
Anritsu ML2438
or equivalent
Instrumentation Accuracy ≤ 0.5 %
Reference Power Sensor
(Cal. Factor and Linearity Tests)
Anritsu MA24002A
or equivalent
NIST Calibration or equivalent
10 dB K Attenuator
(Linearity Test)
Anritsu 41KC-10
VSWR ≤ 1.15, 10 MHz to 6 GHz
6 dB K Attenuator
(Cal. Factor Test)
Anritsu 41KC-6
VSWR ≤ 1.15, 10 MHz to 6 GHz
Adapter N(f) to K(f)
(Cal. Factor and Linearity Tests)
Anritsu 34ANF50 and 34AS50
VSWR ≤ 1.05, 10 MHz to 6 GHz
Power Splitter
(Linearity Tests)
Anritsu K241B
Effective Output VSWR < 1.45, 10 MHz to 6 GHz
Personal Computer
Any