Equipment Description | Manufacturer and Model | Critical Specifications |
Vector Network Analyzer (Pretest) | Anritsu 37369D Anritsu MS4642B or MS4644B (for MA24126A) or equivalent | Reflection Coefficient Uncertainty ≤ 0.013, 10 MHz to 2 GHz Uncertainty ≤ 0.020, 2 GHz to 18 GHz Uncertainty ≤ 0.025, 18 GHz to 26 GHz |
Synthesizer (Freq Response and Linearity Tests) | Anritsu MG3693C or equivalent | Output Power: > +10 dBm, 50 MHz to 26 GHz Output Power Setting Resolution: 0.01 dBm Harmonics: ≤ –40 dBc Source VSWR ≤ 2.00 |
Reference Power Meter (Freq Response and Linearity Tests) | Anritsu ML2438A or equivalent | Instrumentation Accuracy ≤ 0.5 % |
Reference Power Sensor (Freq Response and Linearity Tests) | Anritsu MA24002A or equivalent, or Anritsu MA24004A or equivalent (for MA24126A) | NIST Calibration or equivalent |
10 dB K Attenuator (Linearity Test) | Anritsu 41KC-10 | VSWR ≤ 1.15, 10 MHz to 12 GHz VSWR ≤ 1.20, 12 GHz to 18 GHz VSWR ≤ 1.30, 18 GHz to 26 GHz |
6 dB K Attenuator (Cal. Factor Test) | Anritsu 41KC-6 | VSWR ≤ 1.15, 10 MHz to 12 GHz VSWR ≤ 1.20, 12 GHz to 18 GHz VSWR ≤ 1.30, 18 GHz to 26 GHz |
Adapter N(f) to K(f) (Freq Response and Linearity Tests) | Anritsu 34ANF50 and 34AK50 | VSWR ≤ 1.10, 10 MHz to 18 GHz VSWR ≤ 1.17, 18 GHz to 26 GHz |
Power Divider (Linearity Tests) | Anritsu K241B | Effective Output VSWR < 1.45, 10 MHz to 26 GHz |
Personal Computer | Any |