Refer to the label on the USB sensor for information on frequency range and dynamic range.
Connection
External USB Sensor Connection Setups
Connection and Offset
1. Connect the source to be measured to the USB sensor. Use any required external attenuation or gain so the expected power level is within specification for the sensor.
2. Press Amplitude then Offset Value. Enter an offset value for any external attenuation (negative value) or external gain (positive value). The displayed power will be adjusted by the offset value.
As an example, a power source around 1 Watt (30 dBm) may cause damage to some sensors. 10 dB of in-line external attenuation will bring the power level down to approximately 100 mW, within the sensor’s range. Adding a Offset Value of -10 dB will remove the external attenuation from the displayed power level.