Reflection Calc in Time One Way Round Trip Press this submenu key to toggle the setting for the calculation method that is used in the time domain (not distance) reflection measurements. The method can be either One Way (divide the total time by 2) or Round Trip. Gate Coupled On Off Press this key to toggle the gate coupling On or Off. When set to On, the gate settings (start, stop, and so forth) for all traces are the same. When set to Off, each trace has a gate whose settings are independent of any of the other gates that are being used. Domain Processing Auto BP Only Press this key to toggle domain processing to Auto or to BP Only. In Auto, the instrument selects Low Pass processing whenever possible and switches to Band Pass if not possible. In BP Only, the setting will never be Low Pass. Back Press this key to return to the Application Options Menu (VNA Mode). |