VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :CALCulate{1-16}:APPLication:MEASurement Subsystem
 
:CALCulate{1-16}:APPLication:MEASurement Subsystem
:CALCulate{1-16}:APPLication:MEASurement:TYPe <char>
:CALCulate{1-16}:APPLication:MEASurement:TYPe?
Description
Sets the application type on the indicated channel.
Returns the current application type for the indicated channel where the application types are:
TRANrefl = Transmission/Reflection
NFIGure = Noise Figure
PULSe = PulseView™
Changing to Noise Figure Type
When the VNA application mode is changed to Noise Figure type, the instrument configuration changes to the default Noise Figure trace configuration with:
A single trace
Noise Figure response in a LogMag display type
Changing to Transmission/Reflection or Pulse Type
When the VNA application mode is changed to Transmission/Reflection or Pulse type, the instrument resets to factory as-shipped default configuration setting of four traces set as:
S11 response in a Smith chart display type
S12 response in a LogMag + Phase display type
S21 response in a LogMag + Phase display type
S22 response in a Smith chart display type
Returning to Another User-Defined Preset Configuration
If a different instrument configuration is required programmatically, use the :SYStem:PRESet command subsystem to apply a previously user-defined configuration. The commands are described in the :SYStem subsystem described below. The available commands are:
:SYSTem:PRESet
:SYSTem:PRESet:FILe <string> = 'C:\directory\filename.cha'
:SYSTem:PRESet:FILe?
:SYSTem:PRESet:TYPe <char> = RESET | USER
:SYSTem:PRESet:TYPe?
:SYSTem:PRESet:ZERo
Cmd Parameters
<char> TRANrefl | NFIGure | PULSe
Query Parameters
NA
Output:
<char> TRAN | NFIG | PULS
Range
NA
Default Value
TRANrefl
Syntax Example
:CALC1:APPL:MEAS:TYP NFIG
:CALC1:APPL:MEAS:TYP?