VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :SENSe{1-16}:RECEiver Subsystem
 
:SENSe{1-16}:RECEiver Subsystem
The :SENSe{1-16}:RECEiver subsystem commands are used to configure the VNA receiver.
 
:SENSe{1-16}:RECeiver:CALibration:EXISt? <char>
Description
Query only.
Checks for the existence of a particular receiver calibration on the indicated channel.
Cmd Parameters
NA
Query Parameters
<char> A1 | A2 | A3 | A4 | B1 | B2 | B3 | B4
Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:REC:CAL:EXIS? A1
:SENSe{1-16}:RECeiver:CALibration:STATe <char1>,<char2>
:SENSe{1-16}:RECeiver:CALibration:STATe?
Description
Turns the receiver calibration on/off on the indicated channel.
Returns the receiver calibration on/off status on the indicated channel.
Cmd Parameters
<char1> A1 | A2 | A3 | A4 | B1 | B2 | B3 | B4 | ALL
<char2> 1 | 0 | ON | OFF
Query Parameters
<char1> A1 | A2 | A3 | A4 | B1 | B2 | B3 | B4
Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:REC:CAL:STAT A1,ON
:SENS1:REC:CAL:STAT?A1
 
:SENSe{1-16}:RECeiver:CALibration? <char1>,<char2>
Description
Query only.
Performs the indicated receiver calibration on the indicated channel and outputs status.
Cmd Parameters
NA
Query Parameters
<char1> A1 | A2 | A3 | A4 | B1 | B2 | B3 | B4
<char2> PORT1 | PORT2 | PORT3 | PORT4
Output
<NR1 Data> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:REC:CAL? A1,PORT1
:SENSe{1-16}:RECEiver:CONFiguration <char>
:SENSe{1-16}:RECEiver:CONFiguration?
Description
Sets the receiver configuration on the indicated channel.
The command parameters available depend on the instrument series and model number, where:
• STANdard = The receiver frequency follows the source frequency.
• MSOURce = Multiple source receiver configuration. This can be used with or without actual multiple sources. If selected, allows the receiver frequency to be changed by an equation.
• MMWBBand = Millimeter-Wave broadband receiver configuration. Sets the receiver to be controlled by equations on the VNA appropriate for Broadband/ Millimeter-Wave systems.
• MODBBand = Modular Broadband receiver configuration. Only available on MS4647B VNAs equipped with Broadband 8x option.
MSOURce
For MSOURce (Multiple Source), the equation used depends on whether CW is OFF or ON.
• If CW is OFF, Source = (Multiplier / Divisor) × (Frequency + Offset Frequency)
• If CW is ON, Source = (Multiplier / Divisor) × Offset Frequency
MMWBBand
For MMWBBand (Millimeter-Wave Broadband), the equations used vary for the LO and RF frequencies:
• LO External Source 1 = (Multiplier / Divisor) × (F – 12.35 MHz)
• RF External Source 2 = (Multiplier / Divisor) × F
Returns the receiver configuration on the indicated channel.
Cmd Parameters
The available command and query parameters vary depending on the instrument model and installed options:
• For MS4647B with Option 80/81 = <char> STANdard | MSOURce | MMWBBand | MODBBand
• For MS464xB without Option 80/81 = <char> STANdard | MSOURce | MMWBBand
Query Parameters
<char> STAN | MSOUR | MMWBB | MODBB
Range
NA
Default Value
NA
Syntax Example
:SENS1:REC:CONF STAN
:SENS1:REC:CONF?