VectorStar™ MS464xB Series Microwave Vector Network Analyzer Operation Manual : Initial Checkout Procedures
 
Initial Checkout Procedures
Introduction
The initial checkout procedure chapter describes the detailed procedures for initial operational checkout (such as those performed by incoming quality inspectors) to ensure that the MS464xB Series VNA is fully operational. A quick check procedure is described along with how to determine the instrument’s installed options, current operating system, application software, and firmware versions. Other procedures provided are self-test, non-ratio power, high-level signal, and system noise floor test.
For full performance verification and calibration routines, see the following documents:
VectorStar MS464xB Series VNA Measurement Guide
VectorStar MS464xB Series VNA Maintenance Manual
Warning
This equipment can not be repaired by the operator. DO NOT attempt to remove the equipment covers or to disassemble internal components. Only qualified service technicians with a knowledge of electrical fire and shock hazards should service this equipment. There are high-voltage parts in this equipment presenting a risk of severe injury or fatal electric shock to untrained personnel. In addition, there is a risk of damage to precision components.