PIM Master™ Passive Intermodulation Analyzer with Site Master™ Cable and Antenna Analyzer and 2-Port Option Measurement Guide : PIM versus Time : Amplitude Menu
 
Amplitude Menu
Key Sequence: Amplitude
Amplitude Menu
Reference Level
The reference level is the top graticule line on the display, and can be set from –50 dBm to –160 dBm. A value may be entered from the number keypad. Use the +/– key for negative values. After entering the value press the dBm or dBc submenu key or the Enter key. The Up/Down arrow keys change the reference level in 10 dB steps, and the Left/Right arrow keys change the value by 1 dB.
Scale
The scale can be set in 1 dB steps from 1 dB per division to 15 dB per division. The value can be changed using the number keypad or the Arrow keys.
Auto Range
On Off
Automatically adjusts the reference level based on the input power and y‑axis scale in order to display the trace on screen.
Amplitude Tone
On Off
Toggles On and Off. The frequency of the tone increases as PIM level increases.
Units
dBm dBc
Toggles the measurement display units between dBm and dBc. The amplitude for upper and lower limits automatically adjusts to the equivalent level for the selected units.