PIM Master™ Passive Intermodulation Analyzer with Site Master™ Cable and Antenna Analyzer and 2-Port Option Measurement Guide : PIM versus Time : Setup Menu
 
Setup Menu
Key Sequence: Setup
Setup Menu for PIM vs. Time
High (37–46 dBm) Output Power
Press this submenu key to set the output power level within the higher range for both carrier frequencies, F1 and F2 when a high power test is desired.
Low (20–37 dBm) Output Power
Press this submenu key to set the output power level within the lower range for both carrier frequencies, F1 and F2 when a low power test is desired.
Note: Power output value from the test port is twice the set output value. For example, when Setup > Output Power is set to 20 W, total output power is 40 W.
20 W (F1) + 20 W (F2) = 40 W
Test Duration
Press this submenu key to set the length of time that the test port will deliver power after Measurements > Test is pressed. Range for the test is 1 second to 1200 seconds. Use the Arrow keys or the number keypad. If you use the number keypad, then the Units menu of the Test Duration menu is displayed.
Trace Mode
Fast Low Noise
Press this submenu key to change Trace Mode from the Fast setting to the Low Noise setting. Refer to Trace Mode.
High Reflection Bypass
When on it disables the auto shutoff when a high reflection event occurs.
PIM Aid
Press this submenu key to display the PIM Parameters dialog box (see Figure: PIM Parameters Dialog Box). This dialog box displays and allows editing of the current carrier frequencies and carrier power.