Remote Spectrum Monitor User Guide
:
SCPI SPA Commands
:
:DIAGnostic Subsystem
:DIAGnostic Subsystem
:DIAGnostic:SWEep:TIME?
Title
Measured Sweep Time
Description
This command queries the measured sweep time, in number of milliseconds. This command will return "nan" if no measured sweep time is available, which happens if the sweep was reset and the instrument has not yet swept enough to measure a full sweep.