PowerXpert™ Help : Using the MA243x0A : Uncertainty of a Measurement
 
Uncertainty of a Measurement
Measurement Uncertainty Calculator
Included at the Anritsu download center is a Microsoft Excel tool for calculating power uncertainty. It contains two tabs; one that provides measurement uncertainty for each sensor (selectable from a drop-down menu), and another tab that provides additional uncertainty components and calculated values for the power sensor.
https://www.anritsu.com/en-us/test-measurement/support/downloads/software/dwl003263
Uncertainty Components
Power measurements have many component parts that affect overall measurement uncertainty when measuring power with the sensor:
Measurement Uncertainty: Measurement uncertainty includes the uncertainty associated with the correction of frequency and the linearity response of the sensor over the entire dynamic range. Anritsu follows the industry standard condition of calibrating the power-sensing element at a reference power of 0 dBm (1 mW) and an ambient temperature of 25 °C.
Temperature Compensation: Sensor Temperature Compensation describes the relative power level response over the dynamic range of the sensor. Temperature Compensation should be considered when operating the sensor at other than room temperature.
Noise, Zero Set, and Zero Drift: These are factors within the sensor that impact measurement accuracy at the bottom of the power sensor’s dynamic range.
Mismatch Uncertainty: Mismatch uncertainty is typically the largest component of measurement uncertainty. The error is caused by the differing impedances between the power sensor and the devices to which the power sensor is connected. Mismatch uncertainty can be calculated as follows:
% Mismatch Uncertainty = 100[|1 + Γ1Γ2|2 – 1]
dB Mismatch Uncertainty = 20log|1 + Γ1Γ2|
where:
Γ1 is the reflection coefficient of the power sensor
Γ2 is the reflection coefficient of the device
Uncertainty examples are shown in Table: MA243x0A Measurement Uncertainty Example.
Uncertainty Examples
Two measurement uncertainty calculations are shown for the MA243x0A sensor in Table: MA243x0A Measurement Uncertainty Example. The power sensor is used to measure the power of a 3 GHz, +10.0 dBm and –50 dBm CW signal from a signal source with a 1.3:1 VSWR. The example is based on an aperture time of 20 ms and 64 measurement averages.
MA243x0A Measurement Uncertainty Example 


Uncertainty Term
Uncertainty
Specification
at +10 dBm
(%)
Uncertainty
Specification
at –50 dBm
(%)


Probability Distribution



Divisor
Adjusted Uncertainty
at +10 dBm
(%)
Adjusted Uncertainty
at –50 dBm
(%)
Linearity and Calibration Factor
2.80
2.80
Normal at 2σ
2
1.40
1.40
Noise
0
1.00
Normal at 2σ
2
0.50
0.16
Zero Set
0
0.97
Rectangular
√3
0.56
0.18
Zero Drift
0
0.89
Normal at 2σ
2
0.45
0.14
Mismatch
1.94
1.94
Rectangular
√3
1.12
1.12
Combined Uncertainty (RSS),
30 °C
 
 
 
 
1.79
1.99
Expanded Uncertainty with K=2,
30 °C
 
 
 
 
3.59
3.99
Noise Measurement Uncertainty Calculations for the MA243x0A senors
Noise Calculations at 10 dBm (10 mW):
Noise
1.78 nW/10 mW = 0%
Zero Set
15.6 nW/10 mW = 0%
Zero Drift
17.2 nW/10 mW = 0%
Noise Calculations at –50 dBm (10 nW)
Noise
99.8 pW/10 nW = 1.00%
Zero Set
96.8 pW/10 nW = 0.97%
Zero Drift
89 pW/10 nW = 0.89%