PowerXpert™ Help : Using the MA24105A : Error Messages
 
Error Messages
Error messages are described in Table: Error Messages. The status LED will light amber when an error state occurs. In most cases, the error condition can be easily corrected. If not, note the error message and contact an Anritsu Service Center.
Error Messages
Message
Description
Resolution
Zero invalid as temperature changed by more than 10 Degrees C
The sensor’s ambient temperature has changed by more than 10 ºC since the last zero operation.
Perform the zero operation again.
Temperature out of operating range
The sensor is operating outside of its specified range of 0 ºC to 55 ºC.
Operate the sensor within its specified range.
Sensor zero failed
This message box appears if the zero operation is unsuccessful. The reason could be the presence of RF power at the input of the sensor.
Turn off the RF input to the sensor or disconnect the sensor from the RF source and try the zero operation again.
ZERO_ERROR
This message appears on the application screen if the zero operation is unsuccessful. The reason could be the presence of RF power at the input of the sensor.
Turn off the RF input to the sensor or disconnect the sensor from the RF source and try the zero operation again.
ADC_TEMP_OVERRNGE
This message appears on the application screen if the sensor is being operated in extremely high temperatures and has overheated.
Remove the sensor from the USB connection and allow to cool to the operating range of the sensor:
0 ºC to 55 ºC