VectorStar™ MS464xB Series Microwave Vector Network Analyzer Calibration and Measurement Guide : AutoCal Calibration : Loading a Previously Stored AutoCal Characterization File
 
Loading a Previously Stored AutoCal Characterization File
Use this procedure if the AutoCal Characterization File has already been copied onto the VectorStar hard drive.
Procedure
1. Navigate to CAL KIT/AUTOCAL menu.
MAIN | Calibration | CALIBRATION | Cal Kit/AutoCal Characterization | CAL KIT/AUTOCAL
2. On the CAL KIT/AUTOCAL menu, select Install Kit/Charac.
3. Select the AutoCal Characterization radio button and then click Browse.
4. Navigate to the hard drive (or USB memory device) location of the AutoCal Characterization file for the AutoCal module in use. The recommended directory destination on the VectorStar is C:\Anritsu\VNA\AutoCal
5. Select the AutoCal Characterization file, such as “V123456.ACD,” where “123456” is the serial number of the AutoCal kit. Click Open.
6. Click OK in the Install dialog box. Another install dialog box appears (this one showing the AutoCal controller serial number, which is not the module serial number). Click Install in this box to close it.
The Install dialog box closes and the Cal Kit/AutoCal menu is again available.