This subsystem contains legacy E/O and O/E optical commands. Please see the more current OPTical subsystem which includes these E/O and O/E commands, as well as O/O commands:
Sets the O/E device characterization filename for the multiport E/O measurement on the indicated channel. Query outputs the O/E device characterization filename for the multiport E/O measurement on the indicated channel.
Cmd Parameters
<char> Path and Filename: 'c:\eofiles\myfile1.s2p'
Sets the configuration of the multiport E/O measurement on the indicated channel. Query outputs the configuration of the multiport E/O measurement of the indicated channel.
Cmd Parameters
<char> Command can take any one of the following:
PORT1 | PORT2 | PORT3 | PORT4
Definitions:
PORT1 = Corresponds to Single-ended E/O – Single-ended O/E configuration
PORT2 = Corresponds to Differential E/O – Single-ended O/E configuration
PORT3 = Corresponds to Single-ended E/O – Differential O/E configuration
PORT4 = Corresponds to Differential E/O – Differential O/E configuration
PORT12 = Will result in error “Invalid Configuration”
PORT13 = Will result in error “Invalid Configuration”
PORT14 = Will result in error “Invalid Configuration”
PORT23 = Will result in error “Invalid Configuration”
PORT24 = Will result in error “Invalid Configuration”
Query Parameters
NA
Query Output
<char> PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:EO4M:CONF PORT1
:CALC1:EOOE:EO4M:CONF?
:CALCulate{1-16}:EOOE:EO4Measurment:EOPort <char>
:CALCulate{1-16}:EOOE:EO4Measurment:EOPort?
Description
Sets the configuration of the multiport E/O measurement on the indicated channel. Query outputs the configuration of the multiport E/O measurement of the indicated channel.
Sets the O/E port assignment for the multiport E/O measurement on the indicated channel. Query outputs the O/E port assignment for the multiport E/O measurement of the indicated channel.
De-embed the O/E Device characterization data from the E/O calibration of the indicated channel and return status. This query performs a De-embedding. In order for it to work, the user must first supply O/E device characterization data with an .S2P file and E/O calibration data with a .CHX file.
Sets the calibration filename for the E/O measurement on the indicated channel. Query outputs the calibration filename for the E/O measurement of the indicated channel.
Sets the O/E device characterization filename for the E/O measurement on the indicated channel. Query outputs the O/E device characterization filename for the E/O measurement of the indicated channel.
Sets the O/E device characterization filename for the E/O measurement on the indicated channel. Query outputs the O/E device characterization filename for the E/O measurement of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:EOOE:EOM:CHAR:SWAP ON
:CALC1:EOOE:EOM:CHAR:SWAP?
:CALCulate{1-16}:EOOE:EOMeasurment:EOPort <char>
:CALCulate{1-16}:EOOE:EOMeasurment:EOPort?
Description
Sets the E/O port assignment for the E/O measurement on the indicated channel. Query outputs the E/O port assignment for the E/O measurement of the indicated channel.
Cmd Parameters
<char> PORT1 | PORT2
Query Parameters
NA
Query Output
<char> PORT1 | PORT2
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:EOM:EOP PORT1
:CALC1:EOOE:EOM:EOP?
GO4 Measurement Commands
Note
The commands under this ‘GO4 Measurement’ section are only used for the ‘GoMeasure’ functionality and they do not update the UI.
:CALCulate{1-16}:EOOE:GO4Measurment:CALCulate?
Description
De-embed the O/E Device characterization data from the E/O calibration of the indicated channel and return status.
Sets the O/E calibration filename for the multiport GO measurement on the indicated channel. Query outputs the O/E calibration filename for the multiport GO measurement of the indicated channel.
Sets the O/E device characterization filename for the multiport GO measurement on the indicated channel. Query outputs the O/E device characterization filename for the multiport GO measurement of the indicated channel.
Sets the O/E configuration of the multiport GO measurement on the indicated channel. Query outputs the O/E configuration of the multiport GO measurement of the indicated channel.
Cmd Parameters
<char> Command can take any one of the following:
PORT1 | PORT2 | PORT3 | PORT4
Definitions:
PORT1 = Corresponds to Single-ended E/O – Single-ended O/E configuration
PORT2 = Corresponds to Differential E/O – Single-ended O/E configuration
PORT3 = Corresponds to Single-ended E/O – Differential O/E configuration
PORT4 = Corresponds to Differential E/O – Differential O/E configuration
PORT12 = Will result in error “Invalid Configuration”
PORT13 = Will result in error “Invalid Configuration”
PORT14 = Will result in error “Invalid Configuration”
PORT23 = Will result in error “Invalid Configuration”
PORT24 = Will result in error “Invalid Configuration”
PORT34 – Will result in error “Invalid Configuration”
Query Parameters
NA
Query Output
<char> PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:GO4M:CONF PORT1
:CALC1:EOOE:GO4M:CONF?
:CALCulate{1-16}:EOOE:GO4Measurment:EOPort <char>
:CALCulate{1-16}:EOOE:GO4Measurment:EOPort?
Description
Sets the E/O port assignment for the multiport GO measurement on the indicated channel. Query outputs the E/O port assignment for the multiport GO measurement of the indicated channel.
Sets the O/E port assignment for the multiport GO measurement on the indicated channel. Query outputs the O/E port assignment for the multiport GO measurement of the indicated channel.
Sets the E/O device characterization target filename to use for the multiport GO measurement on the indicated channel. Query outputs the E/O device characterization target filename to use for the multiport GO measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.s2p'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.s2p
Range
N/A
Default Value
N/A
Syntax Example
:CALC1:EOOE:GO4M:TARG 'c:\eofiles\myfile1.s2p'
:CALC1:EOOE:GO4M:TARG?
GO Measurement Commands
Note
The commands under this ‘GO Measurement’ section are only used for the ‘GoMeasure’ functionality and they do not update the UI.
:CALCulate{1-16}:EOOE:GOMeasurment:CALCulate?
Description
Measure and store the E/O device characterization data to the target file of the indicated channel and return status.
Sets the calibration filename for the GO measurement on the indicated channel. Query outputs the calibration filename for the GO measurement of the indicated channel.
Sets the O/E device characterization filename for the GO measurement on the indicated channel. Query outputs the O/E device characterization filename for the GO measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.s2p'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.s2p
Range
N/A
Default Value
N/A
Syntax Example
:CALC1:EOOE:GOM:CHAR 'c:\eofiles\myfile1.s2p'
:CALC1:EOOE:GOM:CHAR?
:CALCulate{1-16}:EOOE:GOMeasurment:EOPort <char>
:CALCulate{1-16}:EOOE:GOMeasurment:EOPort?
Description
Sets the E/O port assignment for the GO measurement on the indicated channel. Query outputs the E/O port assignment for the GO measurement of the indicated channel.
Sets the E/O device characterization target filename to use for the GO measurement on the indicated channel. Query outputs the E/O device characterization target filename to use for the GO measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.s2p'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.s2p
Range
N/A
Default Value
N/A
Syntax Example
:CALC1:EOOE:GOM:TARG 'c:\eofiles\myfile1.s2p'
:CALC1:EOOE:GOM:TARG?
OE4 Measurement Commands
:CALCulate{1-16}:EOOE:OE4Measurment:CALCulate?
Description
De-embed the E/O Device characterization data from the multiport O/E calibration of the indicated channel and return status.
Sets the calibration filename for the multiport O/E measurement on the indicated channel. Query outputs the calibration filename for the multiport O/E measurement of the indicated channel.
Sets the configuration of the multiport O/E measurement on the indicated channel. Query outputs the configuration of the multiport O/E measurement of the indicated channel.
Cmd Parameters
<char> Command can take any one of the following:
PORT1 | PORT2 | PORT3 | PORT4
Definitions:
PORT1 = Corresponds to Single-ended E/O – Single-ended O/E configuration
PORT2 = Corresponds to Differential E/O – Single-ended O/E configuration
PORT3 = Corresponds to Single-ended E/O – Differential O/E configuration
PORT4 = Corresponds to Differential E/O – Differential O/E configuration
PORT12 = Will result in error “Invalid Configuration”
PORT13 = Will result in error “Invalid Configuration”
PORT14 = Will result in error “Invalid Configuration”
PORT23 = Will result in error “Invalid Configuration”
PORT24 = Will result in error “Invalid Configuration”
PORT34 – Will result in error “Invalid Configuration”
Query Parameters
NA
Query Output
<char> PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:OE4M:CONF PORT1
:CALC1:EOOE:OE4M:CONF?
:CALCulate{1-16}:EOOE:OE4Measurment:EOPort <char>
:CALCulate{1-16}:EOOE:OE4Measurment:EOPort?
Description
Sets the E/O port assignment for the multiport O/E measurement on the indicated channel. Query outputs the E/O port assignment for the multiport O/E measurement of the indicated channel.
Sets the O/E port assignment for the multiport O/E measurement on the indicated channel. Query outputs the O/E port assignment for the multiport O/E measurement of the indicated channel.
Sets the calibration filename for the O/E measurement on the indicated channel. Query outputs the calibration filename for the O/E measurement of the indicated channel.
Sets the E/O device characterization filename to use for the O/E measurement on the indicated channel. Query outputs the E/O device characterization filename to use for the O/E measurement of the indicated channel.
Sets the E/O device characterization Swap flag for the O/E measurement on the indicated channel. Query outputs the E/O device characterization Swap flag for the O/E measurement on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:EOOE:OEM:CHAR:SWAP ON
:CALC1:EOOE:OEM:CHAR:SWAP?
:CALCulate{1-16}:EOOE:OEMeasurment:EOPort <char>
:CALCulate{1-16}:EOOE:OEMeasurment:EOPort?
Description
Sets the E/O port assignment for the O/E measurement on the indicated channel. Query outputs the E/O port assignment for the O/E measurement on the indicated channel.