VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :CALCulate{1-16}:EOOE: Subsystem
 
:CALCulate{1-16}:EOOE: Subsystem
Note 
This subsystem contains legacy E/O and O/E optical commands. Please see the more current OPTical subsystem which includes these E/O and O/E commands, as well as O/O commands:
The EOOE subsystem is organized in the following sections:
EOOE Commands – General
EO4 Measurement Commands
E/O Measurement Commands
GO4 Measurement Commands
GO Measurement Commands
OE4 Measurement Commands
O/E Measurement Commands
EOOE Commands – General
 
:CALCulate{1-16}:EOOE:MSGS:LIST?
Description
Outputs a copy of the EOOE messages list.
Cmd Parameters
NA
Query Parameters
NA
Query Output
<arbitrary block data>
Query returns the following list:
0 – Valid
1 – Invalid
2 – InvalidSnPFile
3 – InvalidCHXFile
4 – InvalidCalType
5 – IncompatibleFreq
6 – IncompatiblePort
7 – NoCalExist
8 – WARNING:Extrapolation
9 – InvalidPortSelection
10 – InvalidSnPFileType
11 – InvalidEOFileType
12 – InvalidOEFileType
13 – WARNING:CHXFileContainsOpticalSetup
14 – InvalidPortAssignment
15 – InvalidPortReassignment
16 – NoCHXFileExist
17 – NoSnPFileExist
Range
N/A
Default Value
NA
Syntax Example
:CALC1:EOOE:MSGS:LIST?
 
:CALCulate{1-16}:EOOE:TYPE?
Description
Query only. Outputs the type of E/O-O/E measurement that is currently loaded on the instrument.
Cmd Parameters
NA
Query Parameters
NA
Query Output
<char> EO | OE | OO | NONE
Range
NA
Default Value
NA
Syntax Example
:CALC1:EOOE:TYPE?
EO4 Measurement Commands
 
:CALCulate{1-16}:EOOE:EO4Measurment:CALCulate?
Description
De-embed the O/E Device characterization data from the multiport E/O calibration of the indicated channel and return status.
Cmd Parameters
NA
Query Parameters
NA
Query Output
<NR1> Query returns one of the following:
0 – Valid
1 – Invalid
2 – InvalidSnPFile
3 – InvalidCHXFile
4 – InvalidCalType
5 – IncompatibleFreq
6 – IncompatiblePort
7 – NoCalExist
8 – WARNING:Extrapolation
9 – InvalidPortSelection
10 – InvalidSnPFileType
11 – InvalidEOFileType
12 – InvalidOEFileType
13 – WARNING:CHXFileContainsOpticalSetup
14 – InvalidPortAssignment
15 – InvalidPortReassignment
16 – NoCHXFileExist
17 – NoSnPFileExist
Range
NA
Default Value
NA
Syntax Example
:CALC1:EOOE:EO4M:CALC?
:CALCulate{1-16}:EOOE:EO4Measurment:CHARfile <char>
:CALCulate{1-16}:EOOE:EO4Measurment:CHARfile?
Description
Sets the O/E device characterization filename for the multiport E/O measurement on the indicated channel. Query outputs the O/E device characterization filename for the multiport E/O measurement on the indicated channel.
Cmd Parameters
<char> Path and Filename: 'c:\eofiles\myfile1.s2p'
Query Parameters
NA
Query Output
<char> Query Returns: c:\eofiles\myfile1.s2p
Range
NA
Default Value
NA
Syntax Example
:CALC1:EOOE:EO4M:CHAR “c:\eofiles\myfile1.s2p”
:CALC1:EOOE:EO4M:CHAR?
:CALCulate{1-16}:EOOE:EO4Measurment:CHARfile:SWAP[:STATe] <char>
:CALCulate{1-16}:EOOE:EO4Measurment:CHARfile:SWAP[:STATe]?
Description
Sets the O/E device characterization Swap flag for the multiport E/O measurement on the indicated channel.
The query outputs the O/E device characterization Swap flag for the multiport E/O measurement of the indicated channel.
This command is only valid for Single-ended O/E configurations.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:EOOE:EO4M:CHAR:SWAP ON
:CALC1:EOOE:EO4M:CHAR:SWAP?
:CALCulate{1-16}:EOOE:EO4Measurment:CONFiguration <char>
:CALCulate{1-16}:EOOE:EO4Measurment:CONFiguration?
Description
Sets the configuration of the multiport E/O measurement on the indicated channel. Query outputs the configuration of the multiport E/O measurement of the indicated channel.
Cmd Parameters
<char> Command can take any one of the following:
PORT1 | PORT2 | PORT3 | PORT4
Definitions:
PORT1 = Corresponds to Single-ended E/O – Single-ended O/E configuration
PORT2 = Corresponds to Differential E/O – Single-ended O/E configuration
PORT3 = Corresponds to Single-ended E/O – Differential O/E configuration
PORT4 = Corresponds to Differential E/O – Differential O/E configuration
PORT12 = Will result in error “Invalid Configuration”
PORT13 = Will result in error “Invalid Configuration”
PORT14 = Will result in error “Invalid Configuration”
PORT23 = Will result in error “Invalid Configuration”
PORT24 = Will result in error “Invalid Configuration”
Query Parameters
NA
Query Output
<char> PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:EO4M:CONF PORT1
:CALC1:EOOE:EO4M:CONF?
:CALCulate{1-16}:EOOE:EO4Measurment:EOPort <char>
:CALCulate{1-16}:EOOE:EO4Measurment:EOPort?
Description
Sets the configuration of the multiport E/O measurement on the indicated channel. Query outputs the configuration of the multiport E/O measurement of the indicated channel.
Cmd Parameters
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Query Parameters
NA
Query Output
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:EO4M:EOP PORT12
:CALC1:EOOE:EO4M:EOP?
:CALCulate{1-16}:EOOE:EO4Measurment:OEPort <char>
:CALCulate{1-16}:EOOE:EO4Measurment:OEPort?
Description
Sets the O/E port assignment for the multiport E/O measurement on the indicated channel. Query outputs the O/E port assignment for the multiport E/O measurement of the indicated channel.
Cmd Parameters
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Query Parameters
NA
Query Output
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT2
Syntax Example
:CALC1:EOOE:EO4M:OEP PORT12
:CALC1:EOOE:EO4M:OEP?
E/O Measurement Commands
 
:CALCulate{1-16}:EOOE:EOMeasurment:CALCulate?
Description
De-embed the O/E Device characterization data from the E/O calibration of the indicated channel and return status. This query performs a De-embedding. In order for it to work, the user must first supply O/E device characterization data with an .S2P file and E/O calibration data with a .CHX file.
Cmd Parameters
NA
Query Parameters
NA
Query Output
<NR1> Query returns one of the following:
0 – Valid
1 – Invalid
2 – InvalidSnPFile
3 – InvalidCHXFile
4 – InvalidCalType
5 – IncompatibleFreq
6 – IncompatiblePort
7 – NoCalExist
8 – WARNING:Extrapolation
9 – InvalidPortSelection
10 – InvalidSnPFileType
11 – InvalidEOFileType
12 – InvalidOEFileType
13 – WARNING:CHXFileContainsOpticalSetup
14 – InvalidPortAssignment
15 – InvalidPortReassignment
16 – NoCHXFileExist
17 – NoSnPFileExist
Range
NA
Default Value
NA
Syntax Example
:CALC1:EOOE:EOM:CALC?
:CALCulate{1-16}:EOOE:EOMeasurment:CALFile <string>
:CALCulate{1-16}:EOOE:EOMeasurment:CALFile?
Description
Sets the calibration filename for the E/O measurement on the indicated channel. Query outputs the calibration filename for the E/O measurement of the indicated channel.
Cmd Parameters
<string> Path and file: 'c:\eofiles\myfile1.chx'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.chx
Range
N/A
Default Value
N/A
Syntax Example
:CALC1:EOOE:EOM:CALF 'c:\eofiles\myfile1.chx'
:CALC1:EOOE:EOM:CALF?
:CALCulate{1-16}:EOOE:EOMeasurment:CHARfile <string>
:CALCulate{1-16}:EOOE:EOMeasurment:CHARfile?
Description
Sets the O/E device characterization filename for the E/O measurement on the indicated channel. Query outputs the O/E device characterization filename for the E/O measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.s2p'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.s2p
Range
N/A
Default Value
N/A
Syntax Example
:CALC1:EOOE:EOM:CHAR 'c:\eofiles\myfile1.s2p'
:CALC1:EOOE:EOM:CHAR?
:CALCulate{1-16}:EOOE:EOMeasurment:CHARfile:SWAP[:STATe] <char>
:CALCulate{1-16}:EOOE:EOMeasurment:CHARfile:SWAP[:STATe]?
Description
Sets the O/E device characterization filename for the E/O measurement on the indicated channel. Query outputs the O/E device characterization filename for the E/O measurement of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:EOOE:EOM:CHAR:SWAP ON
:CALC1:EOOE:EOM:CHAR:SWAP?
:CALCulate{1-16}:EOOE:EOMeasurment:EOPort <char>
:CALCulate{1-16}:EOOE:EOMeasurment:EOPort?
Description
Sets the E/O port assignment for the E/O measurement on the indicated channel. Query outputs the E/O port assignment for the E/O measurement of the indicated channel.
Cmd Parameters
<char> PORT1 | PORT2
Query Parameters
NA
Query Output
<char> PORT1 | PORT2
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:EOM:EOP PORT1
:CALC1:EOOE:EOM:EOP?
GO4 Measurement Commands
Note 
The commands under this ‘GO4 Measurement’ section are only used for the ‘GoMeasure’ functionality and they do not update the UI.
:CALCulate{1-16}:EOOE:GO4Measurment:CALCulate?
Description
De-embed the O/E Device characterization data from the E/O calibration of the indicated channel and return status.
Cmd Parameters
NA
Query Parameters
NA
Query Output
<NR1> Query returns one of the following:
0 – Valid
1 – Invalid
2 – InvalidSnPFile
3 – InvalidCHXFile
4 – InvalidCalType
5 – IncompatibleFreq
6 – IncompatiblePort
7 – NoCalExist
8 – WARNING:Extrapolation
9 – InvalidPortSelection
10 – InvalidSnPFileType
11 – InvalidEOFileType
11 – InvalidEOFileType
12 – InvalidOEFileType
13 – WARNING:CHXFileContainsOpticalSetup
14 – InvalidPortAssignment
15 – InvalidPortReassignment
16 – NoCHXFileExist
17 – NoSnPFileExist
Range
Any number from the query parameters list
Default Value
NA
Syntax Example
:CALC1:EOOE:GO4M:CALC?
:CALCulate{1-16}:EOOE:GO4Measurment:CALFile <string>
:CALCulate{1-16}:EOOE:GO4Measurment:CALFile?
Description
Sets the O/E calibration filename for the multiport GO measurement on the indicated channel. Query outputs the O/E calibration filename for the multiport GO measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.chx'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.chx
Range
N/A
Default
N/A
Syntax Example
:CALC1:EOOE:GO4M:CALF 'c:\eofiles\myfile1.chx'
:CALC1:EOOE:GO4M:CALF?
:CALCulate{1-16}:EOOE:GO4Measurment:CHARfile <char>
:CALCulate{1-16}:EOOE:GO4Measurment:CHARfile?
Description
Sets the O/E device characterization filename for the multiport GO measurement on the indicated channel. Query outputs the O/E device characterization filename for the multiport GO measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.s2p'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.s2p
Range
N/A
Default Value
N/A
Syntax Example
:CALC1:EOOE:GO4M:CHAR 'c:\eofiles\myfile1.s2p'
:CALC1:EOOE:GO4M:CHAR?
:CALCulate{1-16}:EOOE:GO4Measurment:CONFiguration <char>
:CALCulate{1-16}:EOOE:GO4Measurment:CONFiguration?
Description
Sets the O/E configuration of the multiport GO measurement on the indicated channel. Query outputs the O/E configuration of the multiport GO measurement of the indicated channel.
Cmd Parameters
<char> Command can take any one of the following:
PORT1 | PORT2 | PORT3 | PORT4
Definitions:
PORT1 = Corresponds to Single-ended E/O – Single-ended O/E configuration
PORT2 = Corresponds to Differential E/O – Single-ended O/E configuration
PORT3 = Corresponds to Single-ended E/O – Differential O/E configuration
PORT4 = Corresponds to Differential E/O – Differential O/E configuration
PORT12 = Will result in error “Invalid Configuration”
PORT13 = Will result in error “Invalid Configuration”
PORT14 = Will result in error “Invalid Configuration”
PORT23 = Will result in error “Invalid Configuration”
PORT24 = Will result in error “Invalid Configuration”
PORT34 – Will result in error “Invalid Configuration”
Query Parameters
NA
Query Output
<char> PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:GO4M:CONF PORT1
:CALC1:EOOE:GO4M:CONF?
:CALCulate{1-16}:EOOE:GO4Measurment:EOPort <char>
:CALCulate{1-16}:EOOE:GO4Measurment:EOPort?
Description
Sets the E/O port assignment for the multiport GO measurement on the indicated channel. Query outputs the E/O port assignment for the multiport GO measurement of the indicated channel.
Cmd Parameters
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Query Parameters
NA
Query Output
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:GO4M:EOP PORT1
:CALC1:EOOE:GO4M:EOP?
:CALCulate{1-16}:EOOE:GO4Measurment:OEPort <char>
:CALCulate{1-16}:EOOE:GO4Measurment:OEPort?
Description
Sets the O/E port assignment for the multiport GO measurement on the indicated channel. Query outputs the O/E port assignment for the multiport GO measurement of the indicated channel.
Cmd Parameters
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Query Parameters
NA
Query Output
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT2
Syntax Example
:CALC1:EOOE:GO4M:EOP Port12
:CALC1:EOOE:GO4M:EOP?
:CALCulate{1-16}:EOOE:GO4Measurment:TARGetfile <string>
:CALCulate{1-16}:EOOE:GO4Measurment:TARGetfile?
Description
Sets the E/O device characterization target filename to use for the multiport GO measurement on the indicated channel. Query outputs the E/O device characterization target filename to use for the multiport GO measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.s2p'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.s2p
Range
N/A
Default Value
N/A
Syntax Example
:CALC1:EOOE:GO4M:TARG 'c:\eofiles\myfile1.s2p'
:CALC1:EOOE:GO4M:TARG?
GO Measurement Commands
Note 
The commands under this ‘GO Measurement’ section are only used for the ‘GoMeasure’ functionality and they do not update the UI.
:CALCulate{1-16}:EOOE:GOMeasurment:CALCulate?
Description
Measure and store the E/O device characterization data to the target file of the indicated channel and return status.
Cmd Parameters
NA
Query Parameters
NA
Query Output
<NR3>
Range
NA
Default Value
NA
Syntax Example
:CALC1:EOOE:GOM:CALC?
:CALCulate{1-16}:EOOE:GOMeasurment:CALFile <string>
:CALCulate{1-16}:EOOE:GOMeasurment:CALFile?
Description
Sets the calibration filename for the GO measurement on the indicated channel. Query outputs the calibration filename for the GO measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.chx'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.chx
Range
N/A
Default
N/A
Syntax Example
:CALC1:EOOE:GOM:CALF 'c:\eofiles\myfile1.chx'
:CALC1:EOOE:GOM:CALF?
:CALCulate{1-16}:EOOE:GOMeasurment:CHARfile <char>
:CALCulate{1-16}:EOOE:GOMeasurment:CHARfile?
Description
Sets the O/E device characterization filename for the GO measurement on the indicated channel. Query outputs the O/E device characterization filename for the GO measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.s2p'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.s2p
Range
N/A
Default Value
N/A
Syntax Example
:CALC1:EOOE:GOM:CHAR 'c:\eofiles\myfile1.s2p'
:CALC1:EOOE:GOM:CHAR?
:CALCulate{1-16}:EOOE:GOMeasurment:EOPort <char>
:CALCulate{1-16}:EOOE:GOMeasurment:EOPort?
Description
Sets the E/O port assignment for the GO measurement on the indicated channel. Query outputs the E/O port assignment for the GO measurement of the indicated channel.
Cmd Parameters
<char> PORT1 | PORT2
Query Parameters
NA
Query Output
<char> PORT1 | PORT2
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:GOM:EOP PORT1
:CALC1:EOOE:GOM:EOP?
:CALCulate{1-16}:EOOE:GOMeasurment:TARGetfile <string>
:CALCulate{1-16}:EOOE:GOMeasurment:TARGetfile?
Description
Sets the E/O device characterization target filename to use for the GO measurement on the indicated channel. Query outputs the E/O device characterization target filename to use for the GO measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.s2p'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.s2p
Range
N/A
Default Value
N/A
Syntax Example
:CALC1:EOOE:GOM:TARG 'c:\eofiles\myfile1.s2p'
:CALC1:EOOE:GOM:TARG?
OE4 Measurement Commands
 
:CALCulate{1-16}:EOOE:OE4Measurment:CALCulate?
Description
De-embed the E/O Device characterization data from the multiport O/E calibration of the indicated channel and return status.
Cmd Parameters
NA
Query Parameters
NA
Query Output
<NR1> Query returns one of the following:
0 – Valid
1 – Invalid
2 – InvalidSnPFile
3 – InvalidCHXFile
4 – InvalidCalType
5 – IncompatibleFreq
6 – IncompatiblePort
7 – NoCalExist
8 – WARNING:Extrapolation
9 – InvalidPortSelection
10 – InvalidSnPFileType
11 – InvalidEOFileType
12 – InvalidOEFileType
13 – WARNING:CHXFileContainsOpticalSetup
14 – InvalidPortAssignment
15 – InvalidPortReassignment
16 – NoCHXFileExist
17 – NoSnPFileExist
Range
NA
Default Value
NA
Syntax Example
:CALC1:EOOE:OE4M:CALC?
:CALCulate{1-16}:EOOE:OE4Measurment:CALFile <string>
:CALCulate{1-16}:EOOE:OE4Measurment:CALFile?
Description
Sets the calibration filename for the multiport O/E measurement on the indicated channel. Query outputs the calibration filename for the multiport O/E measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.chx'
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.chx
Range
N/A
Default Value
N/A
Syntax Example
:CALC1:EOOE:OE4M:CALF 'c:\eofiles\myfile1.chx'
:CALC1:EOOE:OE4M:CALF?
:CALCulate{1-16}:EOOE:OE4Measurment:CHARfile <string>
:CALCulate{1-16}:EOOE:OE4Measurment:CHARfile?
Description
Sets the E/O device characterization filename for the multiport O/E measurement on the indicated channel.
The query outputs the E/O device characterization filename for the multiport O/E measurement of the indicated channel.
Cmd Parameters
<string> “c:\eofiles\myfile1.s2p”
Query Parameters
NA
Query Output
<char> c:\eofiles\myfile1.s2p
Range
N/A
Default Value
N/A
Syntax Example
:CALC1:EOOE:OE4M:CHAR “c:\eofiles\myfile1.s2p”
:CALC1:EOOE:OE4M:CHAR?
:CALCulate{1-16}:EOOE:OE4Measurment:CHARfile:SWAP[:STATe] <char>
:CALCulate{1-16}:EOOE:OE4Measurment:CHARfile:SWAP[:STATe]?
Description
Sets the E/O device characterization Swap flag for the multiport O/E measurement on the indicated channel.
The query outputs the E/O device characterization Swap flag for the multiport O/E measurement of the indicated channel.
This command is only valid for Single-ended E/O configurations.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0 |
Range
NA
Default Value
0
Syntax Example
:CALC1:EOOE:OE4M:CHAR:SWAP ON
:CALC1:EOOE:OE4M:CHAR:SWAP?
:CALCulate{1-16}:EOOE:OE4Measurment:CONFiguration <char>
:CALCulate{1-16}:EOOE:OE4Measurment:CONFiguration?
Description
Sets the configuration of the multiport O/E measurement on the indicated channel. Query outputs the configuration of the multiport O/E measurement of the indicated channel.
Cmd Parameters
<char> Command can take any one of the following:
PORT1 | PORT2 | PORT3 | PORT4
Definitions:
PORT1 = Corresponds to Single-ended E/O – Single-ended O/E configuration
PORT2 = Corresponds to Differential E/O – Single-ended O/E configuration
PORT3 = Corresponds to Single-ended E/O – Differential O/E configuration
PORT4 = Corresponds to Differential E/O – Differential O/E configuration
PORT12 = Will result in error “Invalid Configuration”
PORT13 = Will result in error “Invalid Configuration”
PORT14 = Will result in error “Invalid Configuration”
PORT23 = Will result in error “Invalid Configuration”
PORT24 = Will result in error “Invalid Configuration”
PORT34 – Will result in error “Invalid Configuration”
Query Parameters
NA
Query Output
<char> PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:OE4M:CONF PORT1
:CALC1:EOOE:OE4M:CONF?
:CALCulate{1-16}:EOOE:OE4Measurment:EOPort <char>
:CALCulate{1-16}:EOOE:OE4Measurment:EOPort?
Description
Sets the E/O port assignment for the multiport O/E measurement on the indicated channel. Query outputs the E/O port assignment for the multiport O/E measurement of the indicated channel.
Cmd Parameters
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Query Parameters
NA
Query Output
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:OE4M:EOP PORT12
:CALC1:EOOE:OE4M:EOP?
:CALCulate{1-16}:EOOE:OE4Measurment:OEPort <char>
:CALCulate{1-16}:EOOE:OE4Measurment:OEPort?
Description
Sets the O/E port assignment for the multiport O/E measurement on the indicated channel. Query outputs the O/E port assignment for the multiport O/E measurement of the indicated channel.
Cmd Parameters
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Query Parameters
<char> PORT12 | PORT13 | PORT14 | PORT23 | PORT24 | PORT34 | PORT1 | PORT2 | PORT3 | PORT4
Range
NA
Default Value
PORT2
Syntax Example
:CALC1:EOOE:OE4M:OEP PORT12
:CALC1:EOOE:OE4M:OEP?
O/E Measurement Commands
 
:CALCulate{1-16}:EOOE:OEMeasurment:CALCulate?
Description
De-embed the E/O Device characterization data from the O/E calibration of the indicated channel and return status.
Cmd Parameters
NA
Query Parameters
<NR1> Query returns one of the following:
0 – Valid
1 – Invalid
2 – InvalidSnPFile
3 – InvalidCHXFile
4 – InvalidCalType
5 – IncompatibleFreq
6 – IncompatiblePort
7 – NoCalExist
8 – WARNING:Extrapolation
9 – InvalidPortSelection
10 – InvalidSnPFileType
11 – InvalidEOFileType
12 – InvalidOEFileType
13 – WARNING:CHXFileContainsOpticalSetup
14 – InvalidPortAssignment
15 – InvalidPortReassignment
16 – NoCHXFileExist
17 – NoSnPFileExist
Range
NA
Default Value
NA
Syntax Example
:CALC1:EOOE:OEM:CALC?
:CALCulate{1-16}:EOOE:OEMeasurment:CALFile <string>
:CALCulate{1-16}:EOOE:OEMeasurment:CALFile?
Description
Sets the calibration filename for the O/E measurement on the indicated channel. Query outputs the calibration filename for the O/E measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.chx'
Query Parameters
<char> c:\eofiles\myfile1.chx
Range
N/A
Default
N/A
Syntax Example
:CALC1:EOOE:OEM:CALF 'c:\eofiles\myfile1.chx'
:CALC1:EOOE:OEM:CALF?
:CALCulate{1-16}:EOOE:OEMeasurment:CHARfile <string>
:CALCulate{1-16}:EOOE:OEMeasurment:CHARfile?
Description
Sets the E/O device characterization filename to use for the O/E measurement on the indicated channel. Query outputs the E/O device characterization filename to use for the O/E measurement of the indicated channel.
Cmd Parameters
<string> 'c:\eofiles\myfile1.s2p'
Query Parameters
<char> c:\eofiles\myfile1.s2p
Range
N/A
Default
N/A
Syntax Example
:CALC1:EOOE:OEM:CHAR 'c:\eofiles\myfile1.s2p'
:CALC1:EOOE:OEM:CHAR?
:CALCulate{1-16}:EOOE:OEMeasurment:CHARfile:SWAP[:STATe] <char>
:CALCulate{1-16}:EOOE:OEMeasurment:CHARfile:SWAP[:STATe]?
Description
Sets the E/O device characterization Swap flag for the O/E measurement on the indicated channel. Query outputs the E/O device characterization Swap flag for the O/E measurement on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:EOOE:OEM:CHAR:SWAP ON
:CALC1:EOOE:OEM:CHAR:SWAP?
:CALCulate{1-16}:EOOE:OEMeasurment:EOPort <char>
:CALCulate{1-16}:EOOE:OEMeasurment:EOPort?
Description
Sets the E/O port assignment for the O/E measurement on the indicated channel. Query outputs the E/O port assignment for the O/E measurement on the indicated channel.
Cmd Parameters
<char> PORT1 | PORT2
Query Parameters
<char> PORT1 | PORT2
Range
NA
Default Value
PORT1
Syntax Example
:CALC1:EOOE:OEM:EOP PORT1
:CALC1:EOOE:OEM:EOP?