VectorStar™ MS464xB Series Microwave Vector Network Analyzer User Interface Reference : Advanced Fast CW : Advanced Fast CW Setup and Configuration Menus
 
Advanced Fast CW Setup and Configuration Menus
Fast CW SETUP Menu
Prerequisites
Option 35 and Option 46 is installed on the VNA.
Advanced Fast CW application selected on Application menu.
Previous
Application Menu Set for Advanced Fast CW
Navigation
MAIN | Application | APPLICATION | Advanced Fast CW Setup | FAST CW SETUP
Fast CW SETUP Menu 
Mode
Toggles the mode between Point by Point (PBP) and Full Time Record (FTR). Notes: Depending on the mode selected, the time parameters change from Delay (for PBP) to Start/Stop (for FTR). The bottom trace display status bar indicates the Advanced Fast CW mode (FCW-PBP or FCW-FTR):
Synch Interval
Select activates a toolbar for setting synchronization interval.
Capture Duration
Select activates a toolbar for setting capture duration in time units.
Measurement Width
Select activates a toolbar for setting measurement width in time units.
Delay (Point by Point mode only)
Select activates a toolbar for setting delay in time units.
Start Time (Full Time Record mode only)
Select activates a toolbar for setting start time
Stop Time (Full Time Record mode only)
Select activates a toolbar for setting stop time
# of Points
Select activates a toolbar for setting the number of points measured from Start to Stop time.
Trigger Source
Displays the trigger source that is selected from Other Setup (Addl.FCW Setup) menu. Note that this becomes active as a manual trigger button when Trigger Source is set to Manual.
Additional Fast CW Setup
Select opens the Addl.FCW Setup menu for more Fast CW setup parameters. See Addl. FCW SETUP Menu.
Addl. FCW SETUP Menu
Prerequisites
Option 35 and Option 46 is installed on the VNA.
Advanced Fast CW application selected on Application menu.
Previous
Fast CW SETUP Menu
Navigation
MAIN | Application | APPLICATION | Advanced CW Setup | ADVANCED CW SETUP |Additional Fast CW Setup | ADDL. FCW SETUP
Additional FCW SETUP Menu
Ext Synch Marking
Select opens the External Synch Marking Menu.
Raw Data Export
Select opens the Raw Data Export Menu.
Receiver
Select opens the Receiver Menu.
Rcvr Coupling
Select toggles receiver coupling ON and OFF.
Sampling Interval
Select activates a toolbar for entering desired sampling interval.
Trigger Source
Select opens the Trigger Source Menu.
External Trigger Setup
Select opens the External Trigger Setup Menu.
Manual Trigger Setup
Select opens the Manual Trigger Setup Menu.
GPIB Trigger Setup
Select opens the GPIB Trigger Setup Menu.
External Synch Marking Menu
Prerequisites
Option 35 and Option 46 is installed on the VNA.
Advanced Fast CW application selected on Application menu.
Previous
Addl. FCW SETUP Menu
Navigation
MAIN | Application | APPLICATION | Advanced CW Setup | ADVANCED CW SETUP |Additional Fast CW Setup | ADDL. FCW SETUP | Ext. Synch Marking | EXTERNAL SYNCH
External Synch Marking Menu
OFF
Internal synch is active when OFF is selected.
External Rising
External synch based on the rising edge of the signal present at the Synch In port on the rear panel.
External Falling
External synch based on the falling edge of the signal present at the Synch In port on the rear panel.
Raw Data Export Menu
Prerequisites
Option 35 and Option 46 is installed on the VNA.
Advanced Fast CW application selected on Application menu.
Previous
Addl. FCW SETUP Menu
Navigation
MAIN | Application | APPLICATION | Advanced CW Setup | ADVANCED CW SETUP |Additional Fast CW Setup | ADDL. FCW SETUP | Raw Data Export| RAW DATA EXPORT
Raw Data Export Menu
OFF
Select disables Raw Data Export.
Binary
Select allows data export in binary format. See the Measurement Guide for details on this format.
HDF5 (.mat)
(HDF5 currently not supported; to be supported in later release.)
Binary File Path
Select opens a browsing dialog to designate the file path.
HDF5 File Name
(HDF5 currently not supported; to be supported in later release.)
Receiver Menu
Prerequisites
Option 35 and Option 46 is installed on the VNA.
Advanced Fast CW application selected on Application menu.
Previous
Addl. FCW SETUP Menu
Navigation
MAIN | Application | APPLICATION | Advanced CW Setup | ADVANCED CW SETUP |Additional Fast CW Setup | ADDL. FCW SETUP | Receiver| RECEIVER
Fast CW – Receiver Menu (2-Port Systems)
A1 Receiver
Selects the A1 receiver.
A2 Receiver
Selects the A2 receiver.
B1 Receiver
Selects the B1 receiver.
B2 Receiver
Selects the B2 receiver.
Fast CW – Receiver Menu (4-Port Systems)
A1 Receiver
Selects the A1 receiver.
A2 Receiver
Selects the A2 receiver.
A3 Receiver
Selects the A3 receiver.
A4 Receiver
Selects the A4 receiver.
B1 Receiver
Selects the B1 receiver.
B2 Receiver
Selects the B2 receiver.
B3 Receiver
Selects the B3 receiver.
B4 Receiver
Selects the B4 receiver.
 
Trigger Source Menu
Prerequisites
None (This menu is a duplicate of that from the Sweep/Trigger menus and the entered selections transfer.)
Previous
Addl. FCW SETUP Menu
Navigation
MAIN | Application | APPLICATION | Advanced CW Setup | ADVANCED CW SETUP |Additional Fast CW Setup | ADDL. FCW SETUP | Trigger Source| TRIGGER SOURCE
Fast CW – Trigger Source Menu (2-Port Systems)
Internal
Select sets triggering to be automatically created within the instrument. Internal triggering mode is an automatically triggered point-by-point measurement that is controlled by the instrument internal software.
Manual
Select sets triggering to be manually triggered by the user from the Fast CW SETUP Menu. The default setting is on a per-channel basis. The maximum setting is for all channels.
External
Select sets triggering to be created externally by another instrument and sensed through an external port/connector.
GPIB
Select sets triggering to be through an external GPIB device and communicated to the instrument via a GPIB network.
 
External Trigger Setup Menu
External trigger mode allows a rear panel input to start a measurement based per-point, per-sweep (or per-port), per-channel, or for all channels. The EXT. TRIGGER menu configuration options allow for selection of either a positive or negative signal edge, a trigger delay, and signal handshaking between the external triggering device and the instrument.
Prerequisites
None (This menu is a duplicate of that from the Sweep/Trigger menus and the entered selections transfer.)
Previous
Addl. FCW SETUP Menu
Navigation
MAIN | Application | APPLICATION | Advanced CW Setup | ADVANCED CW SETUP |Additional Fast CW Setup | ADDL. FCW SETUP | External Trigger Setup | EXT. TRIGGER
Fast CW – External Trigger Setup Menu
Trig. All Chan.
The Trigger All Channels button sets the external trigger to measure values for all channels.
Trig. One Chan.
The Trigger One Channel button sets the external trigger to measure values for one channel only.
Trig. One Sweep
The Trigger One Sweep button sets the external trigger to measure values for one sweep only.
Trig. One Point
The Trigger One Point sets the external trigger to measure values for one point only.
Trigger Delay (Time)
Trigger delay sets the time interval between when the instrument received the trigger signal and when the triggered measurement starts.
Select displays the Trigger Delay field toolbar and allows the user to enter a trigger delay time and select units.
Trigger On (Positive Edge/Negative Edge) (External Trigger)
Select toggles the triggering point to be on either the positive edge or negative edge of the triggering signal. The default value is positive edge.
Trigger Handshake (Off/On) (External Trigger)
If triggering handshaking is on, the instrument provides a Ready for Trigger and an Output Trigger signal from the rear panel.
The instrument sends a Ready for Trigger signal to specify if the instrument is ready or not to receive a triggering signal.
The instrument sends an Output Trigger signal when the triggered measurement is complete.
Select toggles whether the trigger handshake is off or on. The default value is OFF.
Manual Trigger Setup Menu
Manual trigger mode is triggered by the user from the Fast CW SETUP Menu to start a measurement based on per-point, per-sweep (or port), per-channel, or for all channels.
Prerequisites
None (This menu is a duplicate of that from the Sweep/Trigger menus and the entered selections transfer.)
Previous
Addl. FCW SETUP Menu
Navigation
MAIN | Application | APPLICATION | Advanced CW Setup | ADVANCED CW SETUP |Additional Fast CW Setup | ADDL. FCW SETUP | Manual Trigger Setup | MANUAL TRIGGER
Fast CW – Manual Trigger Setup Menu
Trig. All Chan.
The Trigger All Channels button sets the trigger to measure values for all channels.
Trig. One Chan.
The Trigger One Channel button sets the trigger to measure value for the active channel only.
Trig. One Sweep
The Trigger One Sweep button sets the trigger to measure values for one sweep only.
Trig. One Point
The Trigger One Point button sets the trigger to measure values for one point only.
 
GPIB Trigger Setup Menu
Prerequisites
None (This menu is a duplicate of that from the Sweep/Trigger menus and the entered selections transfer.)
Previous
Addl. FCW SETUP Menu
Navigation
MAIN | Application | APPLICATION | Advanced CW Setup | ADVANCED CW SETUP |Additional Fast CW Setup | ADDL. FCW SETUP | GPIB Trigger Setup | GPIB TRIGGER
Fast CW – GPIB Trigger Setup Menu
Trig. All Chan.
The Trig. All Chan. button triggers a measurement for all channels.
Trig. One Chan.
The Trig. One Chan. button triggers a measurement for the active channel only.
Trig. One Sweep
The Trig. One Sweep button triggers a measurement for one sweep only.
Trig. One Point
The Trig. One Point button triggers a measurement for one point only.