Noise Floor Time View measurement displays the signal level that is received by the PIM Master at the IM product frequency as a function of time with TX1 and TX2 turned Off.
The Noise Floor Measurement in Figure: Typical Noise Floor (No External Interference, No Limit Line) shows no external interference. A good noise floor that will not interfere with a typical PIM test is below –120 dBm. After verifying that no interference is present, begin PIM testing. Note that units of dBc are not applicable to Noise Floor measurements.
Noise Floor Measurement with No External Interference
Typical Noise Floor (No External Interference, No Limit Line)