PIM Master™ Passive Intermodulation Analyzer with Site Master™ Cable and Antenna Analyzer and 2-Port Option Measurement Guide : Swept PIM : Swept PIM Setup
 
Swept PIM Setup
Note 
The default frequency settings for the PIM Master have been chosen to maximize the swept IM product frequency range within the receive range of the instrument. If required, the default sweep frequencies can be changed in the Swept PIM Aid Menu.
1. Turn on power to the PIM Master.
2. Select the PIM Analyzer mode.
3. Select the Swept PIM measurement from the Measurements menu (refer to Measurements Menu).
4. Connect the test cable between the test port connector of the PIM Master to the device under test (DUT).
Warning 
Confirm that connections are secure. High power RF signals are emitted from the PIM test port and can cause bodily injury. Anritsu recommends using a torque wrench for this connection.
This device (DUT) may be the main feeder cable from the tower or a simple jumper cable. The DUT must be connected to a load device, such as a low PIM load or an Antenna.
Caution 
Do not use a Site Master precision load as the load device because they are not designed to handle the power of the PIM Master and will become damaged immediately.