Symbol Rate Error | After CAL execution at 18 to 28°C, according to the 10 MHz reference. when: Modulation Type: 2FSK, Filter Type: Gaussian, BT=0.5, Symbol Rate 100 ksps, slot length 160 symbol, The signal measured is within the measurement level range and less than or equal to Input Level, and Average = 10 times 30 MHz to 3.5 GHz, Symbol rate error: <±1.0 ppm |
Measurement Level Range | At Pre-Amp Off, or Pre-Amp not installed, refer to your Anritsu receiver or instrument Technical Data Sheet. https://www.anritsu.com/en-US/ |
Transmitter Power Accuracy | After CAL execution at 18 to 28°C, input attenuator ≥ 10 dB, SPAN ≤ 31.25 MHz. The signal measured is within the measurement level range and less than or equal to Input Level - 30 MHz to 3.5 GHz, ± 0.6 dB (at Pre-Amp Off, or Pre-Amp not installed). Transmitter power accuracy is calculated from an RSS (root summed square) error of the absolute amplitude accuracy and the in-band frequency characteristics. |