MX280005A Vector Signal Analysis Software Help : Measurement Ranges and Settings : Introduction : Symbol Rate Error
 
Symbol Rate Error
The Symbol Rate error is listed in Table: Symbol Rate Error
Symbol Rate Error
Symbol Rate Error
After CAL execution at 18 to 28°C, according to the 10 MHz reference. when: Modulation Type: 2FSK, Filter Type: Gaussian, BT=0.5,
Symbol Rate 100 ksps, slot length 160 symbol,
The signal measured is within the measurement level range and less than or equal to Input Level, and Average = 10 times
30 MHz to 3.5 GHz,
Symbol rate error: <±1.0 ppm
Amplitude Measurement
The Amplitude Measurement Specifications are listed in Table: Amplitude Measurement.
Amplitude Measurement
Measurement Level Range
At Pre-Amp Off, or Pre-Amp not installed, refer to your Anritsu receiver or instrument Technical Data Sheet. https://www.anritsu.com/en-US/
Transmitter Power Accuracy
After CAL execution at 18 to 28°C, input attenuator
≥ 10 dB, SPAN ≤ 31.25 MHz. The signal measured is within the measurement level range and less than or equal to Input Level - 30 MHz to 3.5 GHz, ± 0.6 dB (at Pre-Amp Off, or Pre-Amp not installed).
Transmitter power accuracy is calculated from an RSS (root summed square) error of the absolute amplitude accuracy and the in-band frequency characteristics.