MX280005A Vector Signal Analysis Software Help : Measurement Ranges and Settings : Introduction : Modulation/Frequency Measurement
 
Modulation/Frequency Measurement
The Modulation/Frequency Measurement Specifications are listed in Table: Modulation/Frequency Measurement
Modulation/Frequency Measurement
Modulation/Frequency Measurement
Measurement Level Range
At Pre-Amp Off, or Pre-Amp not installed, refer to your Anritsu receiver or instrument Technical Data Sheet. https://www.anritsu.com/en-US/
After CAL performed at 18 to 28°C, For a signal of EVM = 1%, Capture OSR = 4
Carrier Frequency Accuracy
Condition
Carrier Frequency Accuracy
BPSK, QPSK, 8PSK, 16QAM, 32QAM, 64QAM, 128QAM, 256QAM, 2FSK, 4FSK, MSK, π/4DQPSK, 2ASK, 4ASK
Carrier Frequency 30 MHz to 3.5 GHz
(Symbol rate: 4 ksps to 5 Msps)
± (accuracy of reference frequency × carrier frequency + 10 Hz)
 
(512QAM, 1024QAM,
2048QAM)
Carrier Frequency 800 MHz to 3.5 GHz
(Symbol rate: 5 to 50 Msps)
Carrier Frequency 30 MHz to 3.5 GHz
(Symbol rate: 500ksps to 50 Msps)
16APSK, 32APSK)
Carrier Frequency 30 MHz to 3.5 GHz
(Symbol rate: 500ksps to 5 Msps)
Carrier Frequency:
800 MHz to 3.5 GHz
(symbol rate
5 to 50 Msps