PowerXpert™ Help : Verifying MA245xxA Performance : Required Equipment
 
Required Equipment
Required Equipment for MA24507A Power Master is listed in Table: Required Equipment for MA24507A Power Master. Required Equipment for MA24510A Power Master is listed in Table: Required Equipment for MA24510A Power Master.
Required Equipment for MA24507A Power Master
Equipment Description
Manufacturer and Model
Critical Specifications
Vector Network Analyzer
Anritsu
MS4647B (70 GHz)
Reflection Coefficient
UNC 0.012, 50 MHz to 26.5 GHz
UNC 0.018, 26.5 to 40 GHz
UNC 0.026, 40 to 50 GHz
UNC 0.034, 50 to 67 GHz
Synthesizer
Anritsu
MG3697C (70 GHz)
Frequency range 50 MHz to 67 GHz
Output power +5 dBm
Output SWR <2.0
Reference Power Meter
Rhode Schwarz NRP2
Instrumentation Accuracy  0.5 %
Reference Power Sensor
Rhode Schwarz NRP-Z57 (67 GHz)
NIST Calibration or equivalent
10 dB Attenuator
Anritsu
41V-10 (60 GHz)
VSWR ≤1.15, DC to 12 GHz
VSWR ≤1.20, > 12 to 18 GHz
VSWR ≤1.25, > 18 to 26.5 GHz
VSWR ≤1.40, > 26.5 to 40 GHz
VSWR ≤1.90, > 40 to 60 GHz
Personal Computer
Any
Type V Calibration Kit
Anritsu
3654D
 
Required Equipment for MA24510A Power Master
Equipment Description
Manufacturer and Model
Critical Specifications
Vector Network Analyzer
Anritsu
ME7838A (110 GHz)
Reflection Coefficient
UNC 0.015, 50 MHz to 26.5 GHz
UNC 0.016, 26.5 to 40 GHz
UNC 0.017, 40 to 50 GHz
UNC 0.020, 50 to 67 GHz
UNC 0.028 67 to 110 GHz
Synthesizer
Anritsu
MG3697C (70 GHz)
Frequency range 50 MHz to 67 GHz
Output power +5 dBm
Output SWR <2.0
Multiplier (used with synthesizer)
Norden
ND75110G0P0
Keysight E3631A
Flann 27373-WF60
Output Frequency 75 to 110 GHz

12 VDC Output (supply for multiplier)
Adapter inlineWR10 to 1mm female (adapter for multiplier output)
Reference Power Sensor
Rhode Schwarz NRP-Z58 (110 GHz)
NIST Calibration or equivalent
Adapter
Anritsu 34WFV50
For testing MA24510A below 75 GHz.
Type W1 Calibration Kit
Anritsu
3656B
 
Personal Computer
Any