Equipment Description | Manufacturer and Model | Critical Specifications |
Vector Network Analyzer | Anritsu MS4647B (70 GHz) | Reflection Coefficient UNC ≤ 0.012, 50 MHz to 26.5 GHz UNC ≤ 0.018, 26.5 to 40 GHz UNC ≤ 0.026, 40 to 50 GHz UNC ≤ 0.034, 50 to 67 GHz |
Synthesizer | Anritsu MG3697C (70 GHz) | Frequency range 50 MHz to 67 GHz Output power +5 dBm Output SWR <2.0 |
Reference Power Meter | Rhode Schwarz NRP2 | Instrumentation Accuracy ≤ 0.5 % |
Reference Power Sensor | Rhode Schwarz NRP-Z57 (67 GHz) | NIST Calibration or equivalent |
10 dB Attenuator | Anritsu 41V-10 (60 GHz) | VSWR ≤1.15, DC to 12 GHz VSWR ≤1.20, > 12 to 18 GHz VSWR ≤1.25, > 18 to 26.5 GHz VSWR ≤1.40, > 26.5 to 40 GHz VSWR ≤1.90, > 40 to 60 GHz |
Personal Computer | Any | |
Type V Calibration Kit | Anritsu 3654D |
Equipment Description | Manufacturer and Model | Critical Specifications |
Vector Network Analyzer | Anritsu ME7838A (110 GHz) | Reflection Coefficient UNC ≤ 0.015, 50 MHz to 26.5 GHz UNC ≤ 0.016, 26.5 to 40 GHz UNC ≤ 0.017, 40 to 50 GHz UNC ≤ 0.020, 50 to 67 GHz UNC ≤ 0.028 67 to 110 GHz |
Synthesizer | Anritsu MG3697C (70 GHz) | Frequency range 50 MHz to 67 GHz Output power +5 dBm Output SWR <2.0 |
Multiplier (used with synthesizer) | Norden ND75110G0P0 Keysight E3631A Flann 27373-WF60 | Output Frequency 75 to 110 GHz 12 VDC Output (supply for multiplier) Adapter inlineWR10 to 1mm female (adapter for multiplier output) |
Reference Power Sensor | Rhode Schwarz NRP-Z58 (110 GHz) | NIST Calibration or equivalent |
Adapter | Anritsu 34WFV50 | For testing MA24510A below 75 GHz. |
Type W1 Calibration Kit | Anritsu 3656B | |
Personal Computer | Any |