PowerXpert™ Help : Verifying MA245xxA Performance
 
Verifying MA245xxA Performance
 
Introduction
Required Equipment
VSWR Verification
Test Procedure
Frequency Response Test
Introduction
This chapter provides test procedures with measurement recording tables, test equipment setups, and verification tables to verify the measurement accuracy of the MA245xxA Power Master power analyzer. Test and verification is performed by comparing the Power Master to a reference power sensor or Power Master with a specified cal factor and uncertainty. The reference sensor should be calibrated by a reputable standards laboratory using NIST instruments with low published measurement uncertainty values. Sensor test limits in these procedures are set appropriately to the specified comparison equipment. All tests must be performed at an ambient temperature of 20 ºC to 25 ºC to optimize the power sensor’s performance.
Note 
Refer to the instrument Technical Data Sheet for performance specifications.