The :SENSe{1-16}:CORRection:COLLect:TRL:SINGleton subsystem commands provide configuration control of the Thru-Reflect-Line calibration using singletons in different configurations. The commands provide control of singleton configuration parameters and output reporting. The use of the :TRL:SINGleton commands require a 4-port VNA instrument.
TRL Calibration Subsystems
The TRL-related calibration commands are organized into subsystems in the following sections:
Sets the capacitance of the open-like reflection of the TRL singleton calibration on the indicated channel. The C0 (zero) coefficient is measured in Farads
Outputs the capacitance of the open-like reflection of the TRL singleton calibration on the indicated channel.
Sets the C1 (C one) coefficient of the open-like reflection of the TRL singleton calibration on the indicated channel. The C1 coefficient is measured in Farads/Hertz.
Outputs the C1 coefficient of the open-like reflection of the TRL singleton calibration on the indicated channel.
Sets the C2 coefficient of the open-like reflection of the TRL singleton calibration on the indicated channel. The C2 coefficient is measured in Farads/Hertz^2.
Outputs the C2 coefficient of the open-like reflection of the TRL singleton calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2 (Farads per Hertz squared).
Sets the C3 coefficient of the open-like reflection of the TRL singleton calibration on the indicated channel. The C3 coefficient is measured in Farads/Hertz^3.
Outputs the C3 coefficient of the open-like reflection of the TRL singleton calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3 (Farads per Hertz cubed).
Sets the inductance of the TRL singleton short on the indicated channel. The L0 (zero) coefficient is measured in Henrys. Outputs the inductance of the TRL singleton short on the indicated channel.
Sets the L1 (L one) coefficient of the TRL singleton short on the indicated channel. The L1 coefficient is measured in Henrys/Hertz. Outputs the L1 coefficient of the TRL singleton short on the indicated channel.
Sets the L2 coefficient of the TRL singleton short on the indicated channel. The L2 coefficient is measured in Henrys/Hertz^2. Outputs the L2 coefficient of the TRL singleton short on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^2 (Henrys per Hertz squared).
The command sets the L3 coefficient of the TRL singleton short on the indicated channel. The L3 coefficient is measured in Henrys/Hertz^3. The query outputs the L3 coefficient of the TRL singleton short on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^3 (Henrys per Hertz cubed).
The command sets the offset length short-like reflection of the TRL singleton calibration on the indicated channel. The query outputs the offset length of the short-like reflection of the TRL singleton calibration on the indicated channel.