VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :SENSe{1-16}:CORRection:COLLect:TRL[:CALa] Subsystem
 
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa] Subsystem
The:SENSe{1-16}:CORRection:COLLect:TRL[:CALa] subsystem commands control setup and data collection for the TRL calibration. These calibrations are similar to the LRL/LRM family, except one line must be of zero length (hence the term 'thru' in thru-reflect-line).
Calibration Setup Subsystems
These subsystems are used during various phases of calibration configuration setup:
:CALCulate{1-16}:IMPedance:TRANsformation Subsystem
:CALCulate{1-16}:REFerence Subsystem
:SENSe{1-16}:CORRection:COLLect:METHod Subsystem
:SENSe{1-16}:CORRection:COLLect:MICrostrip Subsystem
:SENSe{1-16}:CORRection:COLLect:MSSS Subsystem
:SENSe{1-16}:CORRection:COLLect:MULTiple Subsystem
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:MTRL:CALB Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa] Subsystem (this section)
:SENSe{1-16}:CORRection:COLLect:TRL:CALB Subsystem
:SENSe{1-16}:CORRection:COLLect:TRL:SINGleton Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:TRL:WAVeguide Subsystem
:SENSe{1-16}:CORRection:COLLect Subsystem
:SENSe{1-16}:CORRection:COLLect:WAVeguide Subsystem
:SENSe{1-16}:CORRection:COLLect[:CALa]:PORT Subsystem
:SENSe{1-16}:CORRection:INTerpolation Subsystem
:SENSe{1-16}:CORRection:STATe Subsystem
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND:COUNt <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND:COUNt?
Description
This command sets the number of bands to use in the TRL CALa calibration on the indicated channel.
This query outputs the number of bands to be used the TRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is an integer.
Query Parameters
NA
Query Output
<NR1> The input parameter is an integer.
Range
1 to 5
Default Value
1
Syntax Example
:SENS1:CORR:COLL:TRL:BAND:COUN 5
:SENS1:CORR:COLL:TRL:BAND:COUN?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND:CKIT:LOAD <string>
Description
This command loads a TRL/LRL cal kit file into CALa TRL calibration on the indicated channel.
No query.
Cmd Parameters
<string> Filename and path to the file in the form: 'x:\directory\filename.lcf', where x:\directory\filename.lcf exists and filename.lcf is a valid TRL/LRL cal kit file.
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:TRL:BAND:CKIT:LOAD 'c:\filename.lcf'
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND:CKIT:NAME <string>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND:CKIT:NAME?
Description
This command sets the name that will be stored in the CALa TRL cal kit file on the indicated channel.
Query outputs the name that was last set for the currently loaded CALa TRL cal kit file on the indicated channel.
Cmd Parameters
<string> Filename and path to the file in the form: 'x:\directory\filename.lcf', where x:\directory\filename.lcf exists and filename.lcf is a valid TRL/LRL cal kit file.
Query Parameters
NA
Query Output
<string> The name that is currently associated with the current cal kit file.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:TRL:BAND:CKIT:NAME 'c:\filename.lcf'
:SENS1:CORR:COLL:TRL:BAND:CKIT:NAME?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND:CKIT:SAVE <string>
Description
This command saves a TRL/LRL cal kit file into CALa TRL calibration on the indicated channel.
No query.
Cmd Parameters
<string> Filename and path to the file in the form: 'x:\directory\filename.lcf'
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:TRL:BAND:CKIT:SAVE 'c:\filename.lcf'
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{2-5}:FREQuency:BREakpoint <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{2-5}:FREQuency:BREakpoint?
Description
This command sets the breakpoint frequency between bands of TRL calibration on the indicated channel, where
BAND2: denotes Breakpoint frequency between bands 2 and 1
BAND3: denotes Breakpoint frequency between bands 3 and 2
BAND4: denotes Breakpoint frequency between bands 4 and 3
BAND5: denotes Breakpoint frequency between bands 5 and 4
This query outputs the breakpoint frequency between bands of TRL calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR1> The output parameter is in Hertz.
Range
MPND
Default Value
MPND
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:FREQ:BRE 4E09
:SENS1:CORR:COLL:TRL:BAND2:FREQ:BRE?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:LINE
Description
This command initiates collection of the indicated Band Line Standard data for the CALa calibration on the indicated channel, where
BAND1: denotes the Band1 Cal Device
BAND2: denotes the Band2 Cal Device
BAND3: denotes the Band3 Cal Device
BAND4: denotes the Band4 Cal Device
BAND5: denotes the Band5 Cal Device
No query.
Cmd Parameters
NA
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:TRL:BAND1:LINE
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:REFLection:TYPE <char>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:REFLection:TYPE?
Description
This command sets the reflection type of the TRL CALa calibration on the indicated channel, where:
OPENlike = The reflection type is more like an open.
SHORTlike = The reflection type is more like a short.
This query outputs the reflection type of the TRL CALa calibration on the indicated channel.
Cmd Parameters
<char> OPENlike | SHORTlike
Query Parameters
NA
Query Output
<char> OPEN | SHORT
Range
NA
Default Value
SHORT
Syntax Example
:SENS1:CORR:COLL:TRL:BAND1:REFL:TYPE OPEN
:SENS1:CORR:COLL:TRL:BAND1:REFL:TYPE?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:OPEN:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:OPEN:OFFSet?
Description
Sets the electrical offset length of the open like reflection of the TRL CALa calibration on the indicated channel. Outputs the electrical offset length of the open like reflection of the TRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:OPEN:OFFS 1.0E0
:SENS1:CORR:COLL:TRL:OPEN:OFFS?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:PASSivity:ENForce[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:PASSivity:ENForce[:STATe]?
Description
This command sets the state of whether to use the enforced passivity in the TRL CALa calibration on the indicated channel.
The query outputs whether enforced passivity will be used in the TRL CALa calibration on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:TRL:PASS:ENF 1
:SENS1:CORR:COLL:TRL:PASS:ENF?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:SHORT:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:SHORT:OFFSet?
Description
Sets the offset length of the short like reflection of the TRL CALa calibration on the indicated channel. Outputs the offset length of the short like reflection of the TRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:SHORT:OFFS 1.0E0
:SENS1:CORR:COLL:TRL:SHORT:OFFS?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:LINE:DELay <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:LINE:DELay?
Description
The command sets the delay of the given band for CALa of the TRL calibration of the indicated channel.
The query outputs the delay of the given band for CALa of the TRL calibration of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in seconds.
Query Parameters
NA
Query Output
<NR3> The output parameter is in seconds.
Range
MPND
Default
Default value is dependent on the default dielectric of the substrate and the effective length default value.
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:LINE:DEL 20E-3
:SENS1:CORR:COLL:TRL:BAND2:LINE:DEL?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:LINE:LENGth <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:LINE:LENGth?
Description
Sets the electrical line length of the given band of the TRL CALa calibration on the indicated channel.
Outputs the electrical line length of the given band of the TRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Query Output
NA
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:BAND1:LINE:LENG 1.0e-3
:SENS1:CORR:COLL:TRL:BAND1:LINE:LENG?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:LINE:PLENgth <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:LINE:PLENgth?
Description
The command sets the physical length of the given band for CALa of the TRL calibration of the indicated channel.
The query outputs the physical length of the given band for CALa of the TRL calibration of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in meters.
Query Parameters
NA
Query Output
<NR3> The output parameter is in meters.
Range
MPND
Default
Default value is dependent on the effective length default value.
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:LINE:PLEN 20E-3
:SENS1:CORR:COLL:TRL:BAND2:LINE:PLEN?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH
Description
The command initiates collection of match standard data for the TRL calibration on the indicated channel and port. The use of Port 3 or Port 4 requires a 4-port VNA instrument.
No query.
Cmd Parameters
NA
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:C0 <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:C0?
Description
Sets capacitance of the match device on the indicated channel, band, and port of the TRL CALa calibration. The C0 (C zero) coefficient is in Farads.
Outputs capacitance of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Farads.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Farads.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:BAND1:PORT1:MATCH:C0 3.01E-12
:SENS1:CORR:COLL:TRL:BAND1:PORT1:MATCH:C0?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:C1 <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:C1?
Description
Sets the C1 (C one) coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration. The C1 coefficient is in Farads/Hertz.
Outputs the C1 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz.
Query Parameters
<NR3> The output parameter is in Farads/Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:TRL:BAND1:PORT1:MATCH:C1 2.0E0
:SENS1:CORR:COLL:TRL:BAND1:PORT1:MATCH:C1?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:C2 <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:C2?
Description
Sets the C2 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration. The C2 coefficient is in Farads/Hertz^2.
Outputs the C2 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^2.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:C2 2.0E0
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:C2?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:C3 <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:C3?
Description
Sets the C3 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration. The C3 coefficient is in Farads/Hertz^3.
Outputs the C3 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^3.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:C3 2.0E0
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:C3?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:L0?
Description
Sets inductance of the match device on the indicated channel, band, and port of the TRL CALa calibration. The L0 (zero) coefficient is in Henrys.
Outputs inductance of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
<NR3> The output parameter is in Henrys.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:L0 2.0E-6
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:L0?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:L1?
Description
Sets the L1 (L one) coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration. The L1 coefficient is in Henrys/Hertz.
Outputs the L1 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:L1 1.4
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:L1?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:L2?
Description
Sets the L2 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration. The L2 coefficient is in Henrys/Hertz^2.
Outputs the L2 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^2.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^2.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:L2 2.0E0
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:L2?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:L3?
Description
Sets the L3 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration. The L3 coefficient is in Henrys/Hertz^3.
Outputs the L3 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^3.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^3.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:L3 2.0E0
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:L3?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:OFF1set <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:OFF1set?
Description
Sets the offset Length1 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration. The OFF1 coefficient is measured in Meters/Hertz.
Outputs the offset Length1 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz.
Query Parameters
<NR3> The output parameter is in Meters/Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:OFF1 2.0E0
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:OFF1?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:OFF2set <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:OFF2set?
Description
Sets the offset Length2 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration. The OFF2 coefficient is measured in Meters/Hertz^2.
Outputs the offset Length2 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^2.
Query Parameters
<NR3> The output parameter is in Meters/Hertz^2.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:OFF2 2.0E0
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:OFF2?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:OFF3 <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:OFF3?
Description
Sets the offset Length3 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration. The OFF3 coefficient is measured in Meters/Hertz^3.
Outputs the offset Length3 coefficient of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^3.
Query Parameters
<NR3> The output parameter is in Meters/Hertz^3.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:OFF3 2.0E0
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:OFF3?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:OFFSet?
Description
Sets offset length of the match device on the indicated channel, band, and port of the TRL CALa calibration.
Outputs offset length of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:OFFS 1.0E0
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:OFFS?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:R <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:R?
Description
Sets resistance of the match device on the indicated channel, band, and port of the TRL CALa calibration.
Outputs resistance of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
MPND
Default
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:R 7.5E1
:SENS1:CORR:COLL:TRL:BAND2:PORT1:MATCH:R?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:S1P:FILE <string>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:S1P:FILE?
Description
The command sets the MATCH S1P file on the indicated channel, band, and port for the TRL CALa calibration.
The query outputs the MATCH S1P file on the selected port of the selected device for the TRL CALa calibration for the indicated channel.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<string> s1p file path
Query Parameters
NA
Query Output
<char1> s1p file path
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:TRL:BAND1:PORT1:MATCH:S1P:FILE
'x:\directory\filename.s1p'
:SENS1:CORR:COLL:TRL:BAND1:PORT1:MATCH:S1P:FILE?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:S1P[:STATe]?
Description
Sets the MATCH S1P file settings ON/OFF on the indicated channel, band, and port for the TRL CALa calibration. ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
Outputs the MATCH S1P file settings on the selected port of the indicated channel, band, and port for the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char1>1 | 0
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:TRL:BAND1:PORT1:MATCH:S1P 1
:SENS1:CORR:COLL:TRL:BAND1:PORT1:MATCH:S1P?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:Z0 <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:PORT{1-4}:MATCH:Z0?
Description
Sets impedance of the match device on the indicated channel, band, and port of the TRL CALa calibration.
Outputs impedance of the match device on the indicated channel, band, and port of the TRL CALa calibration.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
MPND
Default
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:TRL:BAND1:PORT1:MATCH:Z0 7.5E1
:SENS1:CORR:COLL:TRL:BAND1:PORT1:MATCH:Z0?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:TYPE <char>
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:BAND{1-5}:TYPE?
Description
The command sets the type of the given band's device in the TRL CALa calibration on the indicated channel.
The query outputs the type of the given band's device in the TRL CALa calibration on the indicated channel.
Cmd Parameters
<char> LINE | MATCH
Query Parameters
NA
Query Output
<char> LINE | MATCH
Range
NA
Default
LINE
Syntax Example
:SENS1:CORR:COLL:TRL:BAND1:TYPE LINE
:SENS1:CORR:COLL:TRL:BAND1:TYPE?
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa]:THRU
Description
The command initiates collection of the THRU data for the CALa calibration on the indicated channel, where the thru length is 0 m.
No query.
Cmd Parameters
NA
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:TRL:THRU