:SENSe{1-16}:NFIGure Subsystem
The :SENSe{1-16}:NFIGure subsystem commands relate to calibrations for noise figure measurements.
Note
These commands can be used for Noise Figure (Option 41) or for Differential Noise Figure (Option 48) for VNA B2 receiver in a 2-port DUT configuration.
Related subsystems for noise figure measurement calibration include:
:SENSe{1-16}:NFIGure:NOISe:CALibration[:STATe]?
Description
For the indicated channel, the query returns the state of whether the Noise Figure noise is applied or not.
Cmd Parameters
NA
Query Parameters
NA
Query Output:
<NR1> 0 | 1
Range
NA
Default
0
Syntax Example
:SENS1:NFIG:NOIS:CAL?
:SENSe{1-16}:NFIGure:LOAD:RCVR:CALibration:DATA <filespec>
Description
For the indicated channel, the command loads the Noise Figure receiver.
Cmd Parameters
Query Parameters
NA
Query Output:
NA
Range
NA
Default
NA
Syntax Example
:SENS1:NFIG:LOAD:RCVR:CAL:DATA ' c:\AnritsuVNA\rcvrfilename.rcvr'
:SENSe{1-16}:NFIGure:RCVR:CALibration:PERForm?
Description
For the indicated channel, the command performs the Noise Figure receiver calibration.
Cmd Parameters
NA
Query Parameters
NA
Query Output:
<NR1> 1 | 0 where 0 = Successful completion of the calibration 1 = Error in the calibration
Range
NA
Default
NA
Syntax Example
:SENS1:NFIG:RCVR:CAL:PERF?
:SENSe{1-16}:NFIGure:RCVR:CALibration[:STATe]?
Description
For the indicated channel, the query returns the state of whether the Noise Figure receiver calibration is applied or not.
Cmd Parameters
NA
Query Parameters
NA
Query Output:
<NR1> 1 | 0
Range
NA
Default
0
Syntax Example
:SENS1:NFIG:RCVR:CAL?
1-800-ANRITSU
10450-00040 Rev. AB