VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :CALCulate{1-16}:PARameter{1-16}:COUPling Subsystem
 
:CALCulate{1-16}:PARameter{1-16}:COUPling Subsystem
The :CALCulate{1-16}:PARameter{TRAC, 1-16}:COUPling subsystem command sets the active trace coupling mode.
Trace Subsystems
Related trace subsystems are:
:CALCulate{1-16}:PARameter and :PARameter{1-16} Subsystem
:CALCulate{1-16}:PARameter{1-16}:FSIMulator Subsystem
:CALCulate{1-16}:PARameter{1-16}:SELect Subsystem
:CALCulate{1-16}:PARameter{1-16}:COUPling Subsystem
:CALCulate{1-16}:POLar Subsystem
:CALCulate{1-16}:PROCessing:ORDer Subsystem
:CALCulate{1-16}[:SELected]:CONVersion Subsystem
:CALCulate{1-16}[:SELected]:DATA Subsystem
:CALCulate{1-16}[:SELected]:FORmat Subsystem
:CALCulate{1-16}[:SELected]:MATH Subsystem
:CALCulate{1-16}[:SELected]:MDATA Subsystem
:CALCulate{1-16}[:SELected]:SMITh Subsystem
:CALCulate{1-16}[:SELected]:SMOothing Subsystem
:CALCulate{1-16}[:SELected]:TRACE{1-16}:MARKer{1-12} Subsystem
:CALCulate{1-16}[:SELected]:TDATA Subsystem
:DISPlay Subsystem
:CALCulate{1-16}:PARameter{1-16}:COUPling <char>
:CALCulate{1-16}:PARameter{1-16}:COUPling?
Description
The command sets the graph format group coupling for the given trace in the indicated channel use. The query outputs the graph format group coupling for the given trace in the indicated channel. The coupling of traces is to facilitate double-y-axis displays. Traces within a coupled group will be forced to the same graph type and the same scale.
Cmd Parameters
<char1> UNCoupled | CPLA | CPLB
Query Parameters
NA
Query Output
<char1> UNC | CPLA | CPLB
Min-Max Range
NA
Default Value
UNC
Syntax Example
:CALC1:PAR1:COUP CPLA
:CALC1:PAR1:COUP?