The :SENSe{1-16}:CORRection:COEFficient subsystem commands are used to simulate a instrument calibration on 2-port VNA instruments so that a set of user-defined calibration parameters can be input using the :SENSe{1-16}:CORRection:COEFficient <char>,<block> command. Note that this subsystem does not perform an actual calibration.
Calibration Simulation Subsystems
These subsystems are used to create a calibrated state in the instrument which is followed by adding the required error correction coefficients for the required calibration type. If this approach is used, each error correction coefficient is entered by separate commands. Simulated calibration subsystems are:
The command simulates a one-path two-port calibration forward direction on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-port VNA or a 4-port VNA instrument. No query. To query the state of this command use:
The command simulates a one-path 2-Port Calibration Reverse Direction on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-port or 4-port VNA instrument. No query. To query the state of this command, use:
The command simulates a full 1-Port Reflection Calibration on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-port VNA or a 4-port VNA instrument. No query. Before sending this command, the simulation port must be specified using:
The command simulates a full 2-Port Calibration on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-port or 4-port VNA instrument. No query. To query the state of this command, use:
The command simulates a full 1-Port Reflection Calibration on both ports on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-port or 4-port VNA instrument. No query. To query the state of this command, use:
The command simulates a 1-Port Response Calibration on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-port or 4-port VNA instrument. No query.
Before sending this command, the simulation port must be specified using:
The command simulates a 1-Port Response Calibration on both ports on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-port or 4-port VNA instrument. No query. To query the state of this command, use:
The command simulates a Transmission Frequency Response Calibration in both directions on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-port or 4-port VNA instrument. No query. To query the state of this command, use:
The command simulates a Transmission Frequency Response Calibration in the forward direction on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-port or 4-port VNA instrument. No query. To query the state of this command, use:
The command simulates a Transmission Frequency Response Calibration in the reverse direction on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-port or 4-port VNA instrument. No query. To query the state of this command, use: