VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :SENSe{1-16}:CORRection:COLLect:LRL:CALB Subsystem
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB Subsystem
The :SENSe{1-16}:CORRection:COLLect:LRL:CALB subsystem commands provide control of Line-Reflect-Line second calibration (or CALB) configuration parameters, execution, and output reporting. The :LRL:CALB commands require a 4-port VNA instrument.
LRL Calibration Subsystems
The LRL-related calibration commands are organized into five (5) subsystems in the following sections:
:SENSe{1-16}:CORRection:COLLect:LRL:BAND{1-5} Subsystem
:SENSe{1-16}:CORRection:COLLect:LRL:CALB Subsystem (this subsystem)
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10} Subsystem
:SENSe{1-16}:CORRection:COLLect:LRL:PORT Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:LRL:SINGleton Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND{1-5}:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND{1-5}:FREQuency?
Description
This command requires a 4-port VNA instrument. The command sets the reference frequency for loss of the given band of the LRL CALB calibration on the indicated channel.
The query outputs the reference frequency for loss of the given band of the LRL CALB calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Hertz.
Range
7E4 to 7E10
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:BAND1:FREQ 1.0E9
:SENS1:CORR:COLL:LRL:CALB:BAND1:FREQ?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND{1-5}:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND{1-5}:LOSS?
Description
This command requires a 4-port VNA instrument. The command sets the line loss of the given band of the LRL CALB calibration on the indicated channel.
The query outputs the line loss of the given band of the LRL CALB calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
NA
Query Output
<NR3> The output parameter is in dB/mm.
Range
7E4 to 7E10
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:BAND1:LOSS 3.0E0
:SENS1:CORR:COLL:LRL:CALB:BAND1:LOSS?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND:COUNt <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND:COUNt?
Description
This command requires a 4-port VNA instrument. The command sets the number of bands to use in the LRL CALB calibration on the indicated channel. The query outputs the number of bands to use in the LRL CALB calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is an integer.
Query Parameters
NA
Query Output
<NR1> The output parameter is an integer.
Range
1 to 5
Default
1
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:BAND:COUN 2
:SENS1:CORR:COLL:LRL:CALB:BAND:COUN?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND{1-5}:REFLection:TYPe <char>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND{1-5}:REFLection:TYPe?
Description
This command requires a 4-port VNA instrument. The command sets the Band reflection type of the LRL CALB calibration on the indicated channel where the type can be set as:
OPENlike = The reflection type is more like an Open.
SHORTlike = The reflection type is more like a Short.
BOTH = The reflection type is like both an Open and a Short.
The query outputs the reflection type of the LRL CALB calibration on the indicated channel for a given band.
Cmd Parameters
<char> OPENlike | SHORTlike | BOTH
Query Parameters
NA
Query Output
<char> OPEN | SHORT | BOTH
Range
NA
Default
OPEN
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:BAND1:REFL:TYP OPENlike
:SENS1:CORR:COLL:LRL:CALB:BAND1:REFL:TYP?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:CKIT:LOAD <string>
Description
This command loads an LRL/TRL cal kit file into CALB LRL calibration on the indicated channel.
Cmd Parameters
<string> Filename and path to the file in the form: 'x:\directory\filename.lcf', where x:\directory\filename.lcf exists and filename.lcf is a valid LRL/TRL cal kit file.
Query Parameters
NA
Query Output
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:CKIT:LOAD 'c:\filename.lcf'
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:CKIT:NAMe <string>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:CKIT:NAMe?
Description
This command sets the file name that will be stored in the LRL/TRL cal kit file.
The query outputs the file name that was last set for the currently loaded file.
Cmd Parameters
<string> Filename and path to the file in the form: 'x:\directory\filename.lcf', where x:\directory\filename.lcf exist and filename.lcf is a valid LRL/TRL cal kit file.
Query Parameters
NA
Query Output
<string> The name that is currently associated with the current cal kit file.
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:CKIT:NAM 'C:\AnritsuVNA\mylrlcalkit.lcf'
:SENS1:CORR:COLL:LRL:CALB:CKIT:NAM?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:CKIT:SAVe <string>
Description
This command saves an LRL cal kit file in the CALB LRL calibration on the indicated channel.
Cmd Parameters
<string> Filename and path to the file in the form: 'x:\directory\filename.lcf'
Query Parameters
NA
Query Output
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:CKIT:SAV 'c:\filename.lcf'
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:LINE
Description
This command requires a 4-port VNA instrument. The command initiates collection of the indicated Device Line Standard data for the CALB calibration on the indicated channel. No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:LINE:DELay <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:LINE:DELay?
Description
This command requires a 4-port VNA instrument. The command sets the delay of the given device for CALB of the LRL calibration of the indicated channel.
The query outputs the delay of the given device for CALB of the LRL calibration of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in seconds.
Query Parameters
NA
Query Output
<NR3> The output parameter is in seconds.
Range
MPND
Default
Default value is dependent on the default dielectric of the substrate and the effective length default value.
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:LINE:DEL 20E-3
:SENS1:CORR:COLL:LRL:CALB:DEV2:LINE:DEL?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:LINE:PLENgth <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:LINE:PLENgth?
Description
This command requires a 4-port VNA instrument. The command sets the physical length of the given device for CALB of the LRL calibration of the indicated channel.
The query outputs the physical length of the given device for CALB of the LRL calibration of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in meters.
Query Parameters
NA
Query Output
<NR3> The output parameter is in meters.
Range
MPND
Default
Default value is dependent on the effective length default value.
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:LINE:PLEN 20E-3
:SENS1:CORR:COLL:LRL:CALB:DEV2:LINE:PLEN?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:LINE:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:LINE:FREQuency?
Description
This command requires a 4-port VNA instrument. The command sets the reference frequency for line loss of the given device on the indicated channel.
The query outputs the reference frequency for loss of the given device of the LRL calibration on the indicated channel.
This command has effect only when an odd-numbered device is entered. Recommend using the following command that sets the Reference Frequency per Band:
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND{1-5}:FREQuency <NRf>
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Hertz.
Range
0 to MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE:FREQ 1.0E7
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE:FREQ?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:LINE:LENGth <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:LINE:LENGth?
Description
This command requires a 4-port VNA instrument. The command sets the line length of the given device of the LRL CALB calibration on the indicated channel.
The query outputs the line length of the given device of the LRL CALB calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Meters.
Range
0 to MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE:LENG 2.5E-2
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE:LENG?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:LINE:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:LINE:LOSS?
Description
This command requires a 4-port VNA instrument. Sets the line loss of the given device of the LRL calibration on the indicated channel.
Outputs the line loss of the given device on the indicated port.
This command has effect only when an odd-numbered device is entered. Recommend using the following command that sets the Loss per Band:
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND{1-5}:LOSS <NRf>
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
NA
Query Output
<NR3> The output parameter is in dB/mm.
Range
NA
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE:LOSS 3.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE:LOSS?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:C0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:C0?
Description
This command requires a 4-port VNA instrument. The command sets the capacitance of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
The query outputs the capacitance of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Farads.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Farads.
Range
MPND
Default
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:C0 3.01E-12
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:C0?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:C1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:C1?
Description
This command requires a 4-port VNA instrument. Sets the C1 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel. Outputs the C1 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Farads/Hertz.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:C1 2.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:C1?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:C2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:C2?
Description
This command requires a 4-port VNA instrument. Sets the C2 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel. Outputs the C2 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Farads/Hertz^2.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:C2 2.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:C2?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:C3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:C3?
Description
This command requires a 4-port VNA instrument. Sets the C3 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel. Outputs the C3 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Farads/Hertz^3.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:C3 2.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:C3?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:L0?
Description
This command requires a 4-port VNA instrument. The command sets the inductance of the match device on the indicated port of the LRL CALB calibration on the indicated channel. Use of Port 3 or Port 4 requires a four-port instrument. The query outputs inductance of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Henrys.
Range
MPND
Default
0
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:L0?
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:L0 2.0E-6
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:L1?
Description
This command requires a 4-port VNA instrument. Sets the L1 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel. Outputs the L1 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Henrys/Hertz.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:L1 1.4
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:L1?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:L2?
Description
This command requires a 4-port VNA instrument. Sets the L2 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel. Outputs the L2 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> in units of Henrys/Hertz^2 (Henrys per Hertz squared)
Query Parameters
NA
Query Output
<NR3> in units of Henrys/Hertz^2
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:L2 10.3E-10
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:L2?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:L3?
Description
This command requires a 4-port VNA instrument. Sets the L3 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel. Outputs the L3 coefficient of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> in Henrys/Hertz^3 (Henrys per Hertz cubed)
Query Parameters
NA
Query Output
<NR3> in Henrys/Hertz^3
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:L3 1.32E-25
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:L3?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:OFFS?
Description
This command requires a 4-port VNA instrument. The command sets the offset length of the match device on the indicated port of the LRL CALB calibration on the indicated channel. The query outputs the offset length of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:OFFS?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:OFF1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:OFF1?
Description
This command requires a 4-port VNA instrument. Sets the offset length coefficient1 of the match device on the indicated port of the LRL CALB calibration on the indicated channel. Outputs the offset length coefficient1 of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Meters/Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:OFF1 1.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:OFF1?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:OFF2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:OFF2?
Description
This command requires a 4-port VNA instrument. Sets the offset length coefficient2 of the match device on the indicated port of the LRL CALB calibration on the indicated channel.Outputs the offset length coefficient2 of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^2.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Meters/Hertz^2.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:OFF2 1.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:OFF2?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:OFF3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:OFF3?
Description
This command requires a 4-port VNA instrument. Sets the offset length coefficient3 of the match device on the indicated port of the LRL CALB calibration on the indicated channel. Outputs the offset length coefficient3 of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^3.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Meters/Hertz^3.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:OFF3 1.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:OFF3?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:S1P:FILE <string>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:S1P:FILE?
Description
This command requires a 4-Port VNA instrument. The command sets the MATCH S1P file on the selected port of the selected device for LRL CALB calibration for the indicated channel. The query outputs the MATCH S1P file on the selected port of the selected device for LRL CALB calibration for the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<string> s1p file path
Query Parameters
NA
Query Output
<char1> s1p file path
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:S1P:FILE
'x:\directory\filename.s1p'
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:S1P:FILE?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:S1P[:STATe]?
Description
This command requires a 4-Port VNA instrument. The command sets the MATCH S1P file settings ON/OFF on the selected port of the selected device for LRL CALB calibration for the indicated channel. ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
The query outputs the MATCH S1P file settings on the selected port of the selected device for LRL CALB calibration for the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default
0
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:S1P 1
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:S1P?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:R <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:R?
Description
This command requires a 4-port VNA instrument. The command sets the resistance of the match device on the indicated port of the LRL CALB calibration on the indicated channel. The query outputs resistance of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Ohms.
Range
MPND
Default
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:R 7.5E1
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:R?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:Z0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:PORT{1-4}:MATCH:Z0?
Description
This command requires a 4-port VNA instrument. The command sets the impedance of the match device on the indicated port of the LRL CALB calibration on the indicated channel. The query outputs the impedance of the match device on the indicated port of the LRL CALB calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Ohms.
Range
MPND
Default
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:Z0 7.5E1
:SENS1:CORR:COLL:LRL:CALB:DEV2:PORT1:MATCH:Z0?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:TYPe <char>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-10}:TYPe?
Description
This command requires a 4-port VNA instrument. The command sets the type of the given device in the LRL CALB calibration on the indicated channel. The query outputs the type of the given device in the LRL CALB calibration on the indicated channel.
Cmd Parameters
<char> LINE | MATCH | DEVICE1 | DEVICE2
Query Parameters
NA
Query Output
<char> LINE | MATCH | DEVICE1 | DEVICE2
Range
NA
Default
LINE
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:TYP LINE
:SENS1:CORR:COLL:LRL:CALB:DEV1:TYP?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:FREQuency:BREakpoint <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:FREQuency:BREakpoint?
Description
This command requires a 4-port VNA instrument. The command sets the breakpoint frequency of the LRL CALB calibration on the indicated channel. The query outputs the breakpoint frequency of the LRL CALB calibration on the indicated channel.
Note: This is a legacy command for breakpoint between Band 1 and Band 2. See the new command:
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:FREQuency:BREakpoint{1-4} <NRf>
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Hertz.
Range
MPND
Default
3.000000000E+009
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:FREQ:BRE 1.0E7
:SENS1:CORR:COLL:LRL:CALB:FREQ:BRE?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:FREQuency:BREakpoint{1-4} <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:FREQuency:BREakpoint{1-4}?
Description
This command requires a 4-port VNA instrument. This command sets the breakpoint frequency between bands of LRL CALB calibration on the indicated channel.
The query outputs the breakpoint frequency between bands of LRL CALB calibration on the indicated channel.
In the command, the number next to BREakpoint has the following correspondence:
BREakpoint1: Breakpoint frequency between bands 2 and 1
BREakpoint2: Breakpoint frequency between bands 3 and 2
BREakpoint3: Breakpoint frequency between bands 4 and 3
BREakpoint4: Breakpoint frequency between bands 5 and 4
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Hertz.
Range
0 to MPND
Default
3E09
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:FREQ:BRE1 4E09
:SENS1:CORR:COLL:LRL:CALB:FREQ:BRE1?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:OPEN:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:OPEN:OFFSet?
Description
This command requires a 4-port VNA instrument. The command sets the electrical offset length of the open like reflection of the LRL CALB calibration on the indicated channel. The query outputs the electrical offset length of the open like reflection of the LRL CALB calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:OPEN:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:CALB:OPEN:OFFS?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:PASSivity:ENForce[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:PASSivity:ENForce[:STATe]?
Description
This command sets the state of whether to use the enforced passivity in the LRL CALB calibration on the indicated channel.
The query outputs whether enforced passivity will be used in the LRL CALB calibration on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:PASS:ENF 1
:SENS1:CORR:COLL:LRL:CALB:PASS:ENF?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:REFPlane <char>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:REFPlane?
Description
This command requires a 4-port VNA instrument. The command sets the location of the reference plane in the LRL CALB calibration on the indicated channel to either the middle or the end of the calibration line. The query outputs the location of the reference plane in the LRL CALB calibration on the indicated channel.
Cmd Parameters
<char> MIDdle | END
Query Parameters
NA
Query Output
<char> MID | END
Range
NA
Default
END
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:REFP MID
:SENS1:CORR:COLL:LRL:CALB:REFP?
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:SHORT:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:SHORT:OFFSet?
Description
This command requires a 4-port VNA instrument. The command sets the offset length of the short like reflection of the LRL CALB calibration on the indicated channel. The query outputs the offset length of the short like reflection of the LRL CALB calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:SHORT:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:CALB:SHORT:OFFS?