VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10} Subsystem
 
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10} Subsystem
The :SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10} subsystem commands provide control of Line-Reflect-Line device parameters, configuration, execution, and output reporting. Some commands require a 4-port VNA instrument.
LRL Calibration Subsystems
The LRL-related calibration commands are organized into five (5) subsystems in the following sections:
:SENSe{1-16}:CORRection:COLLect:LRL:BAND{1-5} Subsystem
:SENSe{1-16}:CORRection:COLLect:LRL:CALB Subsystem
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10} Subsystem
:SENSe{1-16}:CORRection:COLLect:LRL:PORT Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:LRL:SINGleton Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:MATCH:PORT <char>
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:MATCH:PORT?
Description
The command sets the match-port to measure for a match device with both reflection types in the LRL calibration on the indicated channel. The available ports are Port 1 or Port 2. The command can be used by 2-port and 4-port VNA instruments.
The query outputs the match-port to measure for a match device with both reflection types in the LRL calibration on the indicated channel.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
<char> PORT1 | PORT2
Query Parameters
<char> PORT1 | PORT2
Range
NA
Default Value
PORT1
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:MATCH:PORT PORT1
:SENS1:CORR:COLL:LRL:DEV2:MATCH:PORT?
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT{1-4}:MATCH
Description
The command initiates collection of the indicated device match standard data for the LRL calibration on the indicated channel and port. The use of Port 3 or Port 4 requires a 4-port VNA instrument. No query.
This command has effect only when an even-numbered device is entered.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT12:LINE
Description
The command initiates collection of the indicated device line standard data for the LRL calibration on the indicated channel and port-pair. The available port pair is 12 where it is fixed as Port 1 and Port 2. The command can be used by 2-port and 4-port VNA instruments. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT12:LINE
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT12:LINE:DELay <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT12:LINE:DELay?
Description
The command sets the line delay of the given device for LRL calibration on the indicated channel.
The query outputs the line delay of the given device of the LRL calibration on the indicated channel
Cmd Parameters
<NRf> The input parameter is in Seconds.
Query Parameters
<NR3> The output parameter is in Seconds.
Range
MPND
Default Value
Default value is dependent on the default dielectric of the substrate and the effective length default value.
Syntax Example
:SENS1:CORR:COLL:LRL:DEV3:PORT12:LINE:DEL 20E-3
:SENS1:CORR:COLL:LRL:DEV3:PORT12:LINE:DEL?
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT12:LINE:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT12:LINE:FREQuency?
Description
The command sets the reference frequency for loss of the given device on the indicated port-pair of the LRL calibration on the indicated channel. The available port pair is 12 where it is fixed as Port 1 and Port 2. The command can be used by 2-port and 4-port VNA instruments.
The query outputs the reference frequency for loss of the given device on the indicated port-pair of the LRL calibration on the indicated channel. The available port pair is 12.
This command has effect only when an odd-numbered device is entered. Recommend using the following command that sets the Reference Frequency per Band:
:SENSe{1-16}:CORRection:COLLect:LRL:BAND{1-5}:PORT12:FREQuency <NRf>
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Hertz.
Range
0 to MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT12:LINE:FREQ 1.0E9
:SENS1:CORR:COLL:LRL:DEV1:PORT12:LINE:FREQ?
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT12:LINE:LENGth <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT12:LINE:LENGth?
Description
The command sets the effective (air-equivalent) line length of the given device on the indicated port-pair of the LRL calibration on the indicated channel. The available port pair is 12.
The query outputs the effective (air-equivalent) line length of the given device on the indicated port-pair of the LRL calibration on the indicated channel. The available port pair is 12.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Values
Dev X defaults of each band = 0
Dev Y defaults of each band:
Band 1: 5
Band 2: 4
Band 3: 3
Band 4: 2
Band 5: 1
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT12:LINE:LENG 2.0E-3
:SENS1:CORR:COLL:LRL:DEV1:PORT12:LINE:LENG?
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT12:LINE:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT12:LINE:LOSS?
Description
Sets the line loss of the given device on the indicated port-pair of the LRL calibration on the indicated channel. The available port pair is 12.
Outputs the line loss of the given device on the indicated port.
This command has effect only when an odd-numbered device is entered. Recommend using the following command that sets the Loss per Band:
:SENSe{1-16}:CORRection:COLLect:LRL:BAND{1-5}:PORT12:LOSS <NRf>
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
NA
Query Output
<NR3> The output parameter is in dB/mm.
Range
NA
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT12:LINE:LOSS 3.0E0
:SENS1:CORR:COLL:LRL:DEV1:PORT12:LINE:LOSS?
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT12:LINE:PLENgth <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-10}:PORT12:LINE:PLENgth?
Description
The command sets the physical line length of the given device of the LRL calibration on the indicated channel.
The query outputs the physical line length of the given device of the LRL calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
Default value is dependent on effective length default value.
Syntax Example
:SENS1:CORR:COLL:LRL:DEV3:PORT12:LINE:PLEN 20E-3
:SENS1:CORR:COLL:LRL:DEV3:PORT12:LINE:PLEN?