The :CALCulate{1-16}:UFEXtraction subsystem commands provide configuration control and execution for Generalized B (GENB) network extraction functions. The Universal Fixture Extraction option (Option 021) is required for the :CALCulate{1-16}:UFEXtraction commands.
Calibration Option Subsystems
Related calibration option configuration and control subsystems are:
Performs the Network Extraction using Generalized B Method on the indicated channel. Method Type B extracts one 2-port network using a two-tier calibration. Available on 2-port and 4-port VNA instruments.
No query.
The following commands must be sent before performing GenB network extraction:
Sets to use or not use classical open-short in Generalized B Network Extraction on the indicated channel.
Returns the status to use or not use classical open-short in Generalized B on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
1 | 0
Range
NA
Default
0
Syntax Example
:CALC1:UFEX:GENB:OSCL 1
:CALC1:UFEX:GENB:OSCL?
:CALCulate{1-16}:UFEXtraction:GENB:PORT <char>
:CALCulate{1-16}:UFEXtraction:GENB:PORT?
Description
Assigns the port to be used in Generalized B Network Extraction on the indicated channel. The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Returns the selected port to be used in the Generalized B Network Extraction on the indicated channel.
Sets the offset length of the load for the given standard number for Generalized B Network Extraction to be used in Generalized B Network Extraction on the indicated channel
Returns the offset length of the short for the given standard number for Generalized B Network Extraction on the indicated channel.
Sets the reference frequency for the line loss of the open for the given standard number for Generalized B Network Extraction to be used in Generalized B network extraction on the indicated channel.
Returns the reference frequency for the line loss of the open for the given standard number for Generalized B Network Extraction on the indicated channel.
Sets the offset length of the open for the given standard number for Generalized B Network Extraction to be used in Generalized B network extraction on the indicated channel.
Returns the offset length of the open for the given standard number for Generalized B Network Extraction on the indicated channel.
Sets the reference frequency for the line loss of the short for the given standard number to be used in Generalized B network extraction on the indicated channel.
Returns the reference frequency for the line loss of the short for the given standard number for Generalized B Network Extraction on the indicated channel.
Sets the loss of the short for the given standard number for Generalized B Network Extraction to be used in Generalized B network extraction on the indicated channel.
Returns the loss of the short for the given standard number for Generalized B Network Extraction on the indicated channel.
Sets the offset length of the short for the given standard number for Generalized B Network Extraction to be used in Generalized B network extraction on the indicated channel.
Returns the offset length of the short for the given standard number for Generalized B Network Extraction on the indicated channel.
Sets the standard type (short, open, load, or S1P) for the given standard number to be used in Generalized B network extraction on the indicated channel
Returns the standard type for the given standard number for Generalized B Network Extraction on the indicated channel.