VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :SENSe{1-16}:CORRection:COLLect:METHod Subsystem
 
:SENSe{1-16}:CORRection:COLLect:METHod Subsystem
The :SENSe{1-16}:CORRection:COLLect:METHod subsystem command sets the calibration method for the indicated channel. When configuring the instrument calibration this option must be set first, generally followed by line type, and finally calibration type. Optionally, these commands are followed by coefficient and characterization commands for user-defined calibration devices and kits.
Calibration Setup Subsystems
These subsystems are used during various phases of calibration configuration setup:
:CALCulate{1-16}:IMPedance:TRANsformation Subsystem
:CALCulate{1-16}:REFerence Subsystem
:SENSe{1-16}:CORRection:COLLect:METHod Subsystem
:SENSe{1-16}:CORRection:COLLect:MICrostrip Subsystem
:SENSe{1-16}:CORRection:COLLect:MSSS Subsystem
:SENSe{1-16}:CORRection:COLLect:MULTiple Subsystem
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:MTRL:CALB Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:TRL:CALB Subsystem
:SENSe{1-16}:CORRection:COLLect:TRL:SINGleton Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:TRL:WAVeguide Subsystem
:SENSe{1-16}:CORRection:COLLect Subsystem
:SENSe{1-16}:CORRection:COLLect:WAVeguide Subsystem
:SENSe{1-16}:CORRection:COLLect[:CALa]:PORT Subsystem
:SENSe{1-16}:CORRection:INTerpolation Subsystem
:SENSe{1-16}:CORRection:STATe Subsystem
:SENSe{1-16}:CORRection:COLLect:METHod <char>
:SENSe{1-16}:CORRection:COLLect:METHod?
Description
The command sets the calibration method for the indicated channel. The following calibration methods are available:
AUTOcal = Automatic Calibrator (AutoCal) Module calibration method using Anritsu 36585-Series Precision AutoCal Modules
ACLIght = Automatic Calibrator (AutoCal) Module calibration method using Anritsu 36581-Series Standard AutoCal Modules.
LRL = Line-Reflect-Line calibration method
LRM = Line-Reflect-Match calibration method
SOLR = Short-Open-Load-Reciprocal calibration method
SOLT = Short-Open-Load-Through calibration method
SSLT = Short-Short-Load-Through calibration method
SSST = Short-Short-Short-Through calibration method
BBCal = Broadband Cal: Merged Short-Open-Load-Through/Reciprocal (SOLT/R) and Short-Short-Short-Through/Reciprocal (SSST/R)
Note that the type depends on where the breakpoint is set. Sometimes it could be just SOLX, SSST or a merge of both.
MTRL = Multiline Thru-Reflect-Line calibration method
TRX = Thru-Reflect-Line/Match calibration
MSSS = Multiple Offset Short calibration method
The query outputs the calibration method for the indicated channel.
Cmd Parameters
<char> AUTOcal | ACLIght | LRL | LRM | SOLR | SOLT | SSLT | SSST | BBCal | MTRL | TRX | MSSS
Query Output
<char> AUTO | ACLI | LRL | LRM | SOLR | SOLT | SSLT | SSST | BBCal | MTRL | TRX | MSSS
Range
NA
Default Value
SOLT
Syntax Example
:SENS1:CORR:COLL:METH AUTO
:SENS1:CORR:COLL:METH?