VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :SENSe{1-16}:CORRection:COLLect:MTRL[:CALa] Subsystem
 
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa] Subsystem
The:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa] subsystem commands control setup and data collection for the multiline TRL calibration. While similar to the LRL/LRM family, this calibration family can use multiple lines (and a match standard) at every frequency point in a weighted fashion.
Calibration Setup Subsystems
These subsystems are used during various phases of calibration configuration setup:
:CALCulate{1-16}:IMPedance:TRANsformation Subsystem
:CALCulate{1-16}:REFerence Subsystem
:SENSe{1-16}:CORRection:COLLect:METHod Subsystem
:SENSe{1-16}:CORRection:COLLect:MICrostrip Subsystem
:SENSe{1-16}:CORRection:COLLect:MSSS Subsystem
:SENSe{1-16}:CORRection:COLLect:MULTiple Subsystem
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:MTRL:CALB Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:TRL:CALB Subsystem
:SENSe{1-16}:CORRection:COLLect:TRL:SINGleton Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:TRL:WAVeguide Subsystem
:SENSe{1-16}:CORRection:COLLect Subsystem
:SENSe{1-16}:CORRection:COLLect:WAVeguide Subsystem
:SENSe{1-16}:CORRection:COLLect[:CALa]:PORT Subsystem
:SENSe{1-16}:CORRection:INTerpolation Subsystem
:SENSe{1-16}:CORRection:STATe Subsystem
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:REFLection:TYPE <char>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:REFLection:TYPE?
Description
This command sets the reflection type of the mTRL CALa calibration on the indicated channel, where:
OPENlike = The reflection type is more like an open.
SHORTlike = The reflection type is more like a short.
This query outputs the reflection type of the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<char> OPENlike | SHORTlike
Query Parameters
NA
Query Output
<char> OPEN | SHORT
Range
NA
Default Value
SHORT
Syntax Example
:SENS1:CORR:COLL:MTRL:REFL:TYPE OPEN
:SENS1:CORR:COLL:MTRL:REFL:TYPE?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:CKIT:LOAD <string>
Description
This command loads an mTRL cal kit file into CALa mTRL calibration on the indicated channel. No query.
Cmd Parameters
<string> Filename and path to the file in the form: x:\directory\filename.mlcf, where x:\directory\filename.mlcf exists and filename.mlcf is a valid mTRL cal kit file.
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:MTRL:CKIT:LOAD 'c:\filename.mlcf'
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:CKIT:NAME <string>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:CKIT:NAME?
Description
This command sets the name that will be stored in the CALa mTRL cal kit file on the indicated channel.
The query outputs the name that was last set for the currently loaded CALa mTRL cal kit file on the indicated channel.
Cmd Parameters
<string> Filename and path to the file in the form: x:\directory\filename.mlcf, where x:\directory\filename.mlcf exists and filename.mlcf is a valid mTRL cal kit file.
Query Parameters
NA
Query Output
<string> The name that is currently associated with the current cal kit file.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:MTRL:CKIT:NAME 'c:\filename.mlcf'
:SENS1:CORR:COLL:MTRL:CKIT:NAME?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:CKIT:SAVE <string>
Description
This command saves a mTRL cal kit file into CALa mTRL calibration on the indicated channel.
Cmd Parameters
<string> Filename and path to the file in the form: x:\directory\filename.mlcf
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:MTRL:CKIT:SAVE 'c:\filename.mlcf'
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE{1-10}:MEASure
Description
This command initiates collection of the Line Standard data for the indicated line for the mTRL CALa calibration on the indicated channel. No query.
Cmd Parameters
NA
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:MTRL:LINE1:MEAS
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE{1-10}:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE{1-10}:FREQuency?
Description
This command sets the reference frequency on the indicated line of the mTRL CALa calibration on the indicated channel.
The query outputs the reference frequency on the indicated line of the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output is in Hertz.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:LINE1:FREQ 1.0E7
:SENS1:CORR:COLL:MTRL:LINE1:FREQ?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE{1-10}:LENGth <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE{1-10}:LENGth?
Description
This command sets the line length on the indicated line of the mTRL CALa calibration on the indicated channel.
The query outputs the line length on the indicated line of the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output is in Meters.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:LINE1:LENG 1.0E0
:SENS1:CORR:COLL:MTRL:LINE1:LENG?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE{1-10}:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE{1-10}:LOSS?
Description
This command sets the line loss on the indicated line of the mTRL CALa calibration on the indicated channel.
The query outputs the line loss on the indicated line of the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
NA
Query Output
<NR3> The output is in dB/mm.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:LINE1:LOSS 1.0E0
:SENS1:CORR:COLL:MTRL:LINE1:LOSS?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:OPEN:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:OPEN:OFFSet?
Description
This command sets the electrical offset length of the open-like reflection of the mTRL CALa calibration on the indicated channel.
The query outputs the electrical offset length of the open-like reflection of the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output is in Meters.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:OPEN:OFFS 1.0E0
:SENS1:CORR:COLL:MTRL:OPEN:OFFS?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:SHORT:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:SHORT:OFFSet?
Description
This command sets the offset length of the short-like reflection of the mTRL CALa calibration on the indicated channel.
The query outputs the offset length of the short-like reflection of the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output is in Meters.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:SHORT:OFFS 1.0E0
:SENS1:CORR:COLL:MTRL:SHORT:OFFS?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:MEASure
Description
This command initiates collection of match standard data for the mTRL CALa calibration on the indicated channel and port.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
No query.
Cmd Parameters
NA
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:MEAS
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PASSivity:ENForce[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PASSivity:ENForce[:STATe]?
Description
This command sets the state of whether to use the enforced passivity in the mTRL CALa calibration on the indicated channel.
The query outputs whether enforced passivity will be used in the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:MTRL:PASS:ENF 1
:SENS1:CORR:COLL:MTRL:PASS:ENF?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:C0 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:C0?
Description
This command sets capacitance of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The query outputs capacitance of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Farads.
Query Parameters
NA
Query Output
<NR3> The output is in Farads.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:C0 3.01E-12
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:C0?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:C1 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:C1?
Description
This command sets C1 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The query outputs the C1 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz.
Query Parameters
NA
Query Output
<NR3> The output is in Farads/Hertz.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:C1 2.0E0
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:C1?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:C2 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:C2?
Description
This command sets C2 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The query outputs the C2 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2.
Query Parameters
NA
Query Output
<NR3> The output is in Farads/Hertz^2.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:C2 2.0E0
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:C2?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:C3 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:C3?
Description
This command sets C3 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The query outputs the C2 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3.
Query Parameters
NA
Query Output
<NR3> The output is in Farads/Hertz^3.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:C3 2.0E0
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:C3?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:L0?
Description
This command sets inductance of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The query outputs inductance of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
NA
Query Output
<NR3> The output is in Henrys.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:L0 2.01E-6
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:L0?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:L1?
Description
This command sets L1 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The query outputs L1 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
NA
Query Output
<NR3> The output is in Henrys/Hertz.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:L1 2.01E0
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:L1?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:L2?
Description
This command sets L2 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The query outputs L2 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^2.
Query Parameters
NA
Query Output
<NR3> The output is in Henrys/Hertz^2.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:L2 2.01E0
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:L2?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:L3?
Description
This command sets L3 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The query outputs L3 coefficient of match on the indicated port of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^3.
Query Parameters
NA
Query Output
<NR3> The output is in Henrys/Hertz^3.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:L3 2.01E0
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:L3?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:OFFS?
Description
This command sets the offset length of match of the mTRL CALa calibration on the indicated channel.
The query outputs the offset length of match of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:OFFS 1.0E0
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:OFFS?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:OFF1 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:OFF1?
Description
This command sets the offset length1 coefficient of match of the mTRL CALa calibration on the indicated channel.
The query outputs the offset length1 coefficient of match of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz.
Query Parameters
NA
Query Output
<NR3> The output is in Meters/Hertz.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:OFF1 2.0E0
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:OFF1?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:OFF2 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:OFF2?
Description
This command sets the offset length2 coefficient of match of the mTRL CALa calibration on the indicated channel.
The query outputs the offset length2 coefficient of match of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^2.
Query Parameters
NA
Query Output
<NR3> The output is in Meters/Hertz^2.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:OFF2 2.0E0
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:OFF2?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:OFF3 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:OFF3?
Description
This command sets the offset length3 coefficient of match of the mTRL CALa calibration on the indicated channel.
The query outputs the offset length3 coefficient of match of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^3.
Query Parameters
NA
Query Output
<NR3> The output is in Meters/Hertz^3.
Range
MPND
Default Value
00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:OFF3 2.0E0
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:OFF3?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:R <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:R?
Description
This command sets the resistance of match of the mTRL CALa calibration on the indicated channel.
The query outputs the resistance of match of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
NA
Query Output
<NR3> The output is in Ohms.
Range
MPND
Default Value
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:R 7.5E1
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:R?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:Z0 <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:Z0?
Description
This command sets the impedance of match of the mTRL CALa calibration on the indicated channel.
The query outputs the impedance of match of the mTRL CALa calibration on the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
NA
Query Output
<NR3> The output is in Ohms.
Range
MPND
Default Value
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:Z0 7.5E1
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:Z0?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE{1-10}:PLENgth <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE{1-10}:PLENgth?
Description
This command sets the physical line length on the indicated line of the mTRL CALa calibration on the indicated channel.
The query outputs the physical line length on the indicated line of the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output is in Meters.
Range
MPND
Default Value
Default value is dependent on effective length default value.
Syntax Example
:SENS1:CORR:COLL:MTRL:LINE1:PLEN 20E-3
:SENS1:CORR:COLL:MTRL:LINE1:PLEN?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE{1-10}:DELay <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE{1-10}:DELay?
Description
This command sets the line delay on the indicated line of the mTRL CALa calibration on the indicated channel.
The query outputs the line delay on the indicated line of the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Seconds.
Query Parameters
NA
Query Output
<NR3> The output is in Seconds.
Range
MPND
Default Value
Default value is dependent on the default dielectric of the substrate and the effective length default value.
Syntax Example
:SENS1:CORR:COLL:MTRL:LINE1:DEL 20E-3
:SENS1:CORR:COLL:MTRL:LINE1:DEL?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:S1P:FILE <string>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:S1P:FILE?
Description
This command sets the match s1p file on the indicated port of the mTRL CALa calibration on the indicated channel.
The query outputs the match s1p file on the indicated port of the mTRL CALa calibration on the indicated channel.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<string> Filename and path to the file in the form: x:\directory\filename.s1p, where x:\directory\filename.s1p must exist.
Query Parameters
NA
Query Output
<string> Filename and path in the form: x:\directory\filename.s1p.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:S1P:FILE 'c:\filename.s1p'
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:S1P:FILE?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:S1P[:STATe]?
Description
This command sets the MATCH S1P file settings ON/OFF on the selected port for mTRL CALa calibration for the indicated channel. ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
The query outputs the MATCH S1P file settings on the selected port for mTRL CALa calibration for the indicated channel.
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:S1P 1
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:S1P?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:REFPlane <char>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:REFPlane?
Description
This command sets the location of the reference plane in the mTRL CALa calibration on the indicated channel where:
MIDdle = the middle of the transmission line
END = the end of the transmission line
The query outputs the location of the reference plane in the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<char> MIDdle | END
Query Parameters
NA
Query Output
<char> MID | END
Range
NA
Default Value
END
Syntax Example
:SENS1:CORR:COLL:MTRL:REFP MID
:SENS1:CORR:COLL:MTRL:REFP?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE:COUNt <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:LINE:COUNt?
Description
This command sets the number of lines to use in the mTRL CALa calibration on the indicated channel.
The query outputs the number of lines to use in the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is an integer.
Query Parameters
NA
Query Output
<NR1> The output is an integer.
Range
1 to 10
Default Value
2
Syntax Example
:SENS1:CORR:COLL:MTRL:COUN 5
:SENS1:CORR:COLL:MTRL:COUN?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:MATCH[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:MATCH[:STATe]?
Description
This command sets the state of whether to include match in the mTRL CALa calibration on the indicated channel. When enabled, the match standard will be used whenever a line length delta falls below the phase threshold specified in the match definition section.
The query outputs whether match will be included in the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:MTRL:MATCH 1
:SENS1:CORR:COLL:MTRL:MATCH?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:THReshold <double>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:PORT{1-4}:MATCH:THReshold?
Description
This command requires a 4-port VNA instrument. This command the electrical line length delta for match use for CALa calibration on the indicated channel.
The query outputs the electrical line length delta for match use for CALa calibration on the indicated channel.
Cmd Parameters
<double> The input is in degrees.
Query Parameters
NA
Query Output
<string> The output is in degrees.
Range
[0-180]
Default Value
20.0
Syntax Example
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:THR 50
:SENS1:CORR:COLL:MTRL:PORT1:MATCH:THR?
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:OPTLength[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa]:OPTLength[:STATe]?
Description
This command sets the state of whether to use the Optimal Length Super-Weighting in the mTRL CALa calibration on the indicated channel. When enabled, the weighting of the various line measurements will change in computing the calibration error coefficients. The line pair with the optimal length delta for the given frequency point will be weighted more heavily and unweighting due to repeatability differences will be reduced.
The query outputs whether Optimal Length Super-Weighting will be used in the mTRL CALa calibration on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:MTRL:OPTL 1
:SENS1:CORR:COLL:MTRL:OPTL?
:SENSe{1-16}:CORRection:COLLect:MTRL:WAVeguide:DIELectric <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL:WAVeguide:DIELectric?
Description
This command sets the mTRL waveguide calibration dielectric for the indicated channel.
The query outputs the mTRL waveguide calibration dielectric for the indicated channel.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
NA
Query Output
<NR3> The output is a unitless number.
Range
MPND
Default Value
1.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:WAV:DIEL 1.2E0
:SENS1:CORR:COLL:MTRL:WAV:DIEL?
:SENSe{1-16}:CORRection:COLLect:MTRL:WAVeguide:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:MTRL:WAVeguide:FREQuency?
Description
This command sets the mTRL waveguide calibration cutoff frequency for the indicated channel.
The query outputs the mTRL waveguide calibration cutoff frequency for the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output is in Hertz.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MTRL:WAV:FREQ 1E10
:SENS1:CORR:COLL:MTRL:WAV:FREQ?