VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :SENSe{1-16}:CORRection:COLLect Subsystem
 
:SENSe{1-16}:CORRection:COLLect Subsystem
The :SENSe{1-16}:CORRection:COLLect subsystem sets various coefficients and parameters for a pending calibration of the indicated channel.
Calibration Setup Subsystems
These subsystems are used during various phases of calibration configuration setup:
:CALCulate{1-16}:IMPedance:TRANsformation Subsystem
:CALCulate{1-16}:REFerence Subsystem
:SENSe{1-16}:CORRection:COLLect:METHod Subsystem
:SENSe{1-16}:CORRection:COLLect:MICrostrip Subsystem
:SENSe{1-16}:CORRection:COLLect:MSSS Subsystem
:SENSe{1-16}:CORRection:COLLect:MULTiple Subsystem
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:MTRL:CALB Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:TRL:CALB Subsystem
:SENSe{1-16}:CORRection:COLLect:TRL:SINGleton Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:TRL:WAVeguide Subsystem
:SENSe{1-16}:CORRection:COLLect Subsystem
:SENSe{1-16}:CORRection:COLLect:WAVeguide Subsystem
:SENSe{1-16}:CORRection:COLLect[:CALa]:PORT Subsystem
:SENSe{1-16}:CORRection:INTerpolation Subsystem
:SENSe{1-16}:CORRection:STATe Subsystem
:SENSe{1-16}:CORRection:COLLect:REFerence:Z0 <NRf>
:SENSe{1-16}:CORRection:COLLect:REFerence:Z0?
Description
Sets the Z0 (Z zero) reference impedance to use for the calibration on the indicated channel. Outputs the reference impedance to use for the calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
1E-4 to 1E10
Default Value
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:REF:Z0 7.5E1
:SENS1:CORR:COLL:REF:Z0?
:SENSe{1-16}:CORRection:COLLect:SAVe
Description
Initiates calculation of correction coefficients for the programmed calibration type on the indicated channel. This command should only be initiated if all required measurements have been collected. If the command is issued before all measurements are completed, an error is generated. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:SAV
:SENSe{1-16}:CORRection:COLLect:SMCorrection[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:SMCorrection[:STATe]?
Description
The command sets whether the secondary match correction for calibration will be used.
The query outputs whether the secondary match correction for calibration will be used.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:SMC ON
:SENS1:CORR:COLL:SMC?
:SENSe{1-16}:CORRection:COLLect:TFR:CLEar
Description
Clears the list of Transmission Frequency Response calibrations. No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:TFR:CLE
:SENSe{1-16}:CORRection:COLLect:THRu:ADD <char>
Description
Adds Transmission Throughs (Thrus) to the selected 3-Port Calibration or 4-Port Calibration on the indicated channel. The use of Port 3 and/or Port 4 requires a 4-port VNA instrument. No query. The available throughs (thrus) are:
THRu12 = Through line between Port 1 and Port 2; short form is THR12.
THRu13 = Through line between Port 1 and Port 3; short form is THR13.
THRu14 = Through line between Port 1 and Port 4; short form is THR14.
THRu23 = Through line between Port 2 and Port 3; short form is THR23.
THRu24 = Through line between Port 2 and Port 4; short form is THR24.
THRu34 = Through line between Port 3 and Port 4; short form is THR34.
Cmd Parameters
<char> THRu12 | THRu13 | THRu14 | THRu23 | THRu24 | THRu34
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:THR:ADD THRU12
:SENSe{1-16}:CORRection:COLLect:THRu:CLEar
Description
Clears the Transmission Thrus of the selected 3-Port Calibration or 4-Port Calibration on the indicated channel.
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:THR:CLE