The :SENSe{1-16}:CORRection:COLLect:WAVeguide subsystem sets the calibration parameters and coefficients for waveguide line types. Use this subsystem to set waveguide parameters before starting a calibration.
Calibration Setup Subsystems
These subsystems are used during various phases of calibration configuration setup:
Sets the waveguide calibration kit dielectric value on the indicated channel. Outputs the waveguide calibration kit dielectric value on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit cutoff frequency on the indicated channel. Outputs the waveguide calibration kit cutoff frequency on the indicated channel.
Sets the waveguide kit type on the indicated channel. Outputs the waveguide kit type on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit label on the indicated channel. Outputs the waveguide calibration kit label on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
Sets the waveguide calibration kit L0 (L zero) load inductance on the indicated channel. Outputs the waveguide calibration kit load inductance on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit load offset on the indicated channel. Outputs the waveguide calibration kit load offset on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit load resistance on the indicated channel. Outputs the waveguide calibration kit load resistance on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit open capacitance C0 (C zero) value on the indicated channel. The C0 coefficient is measured in Farads. Outputs the waveguide calibration kit open capacitance C0 value on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit open capacitance C1 (C one) term on the indicated channel. The C1 coefficient is measured in Farads/Hertz. Outputs the waveguide calibration kit open capacitance1 term on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit open capacitance C2 term on the indicated channel. The C2 coefficient is measured in Farads/Hertz^2. Outputs the waveguide calibration kit open capacitance C2 term on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit open capacitance C3 term on the indicated channel. The C3 coefficient is measured in Farads/Hertz^3. Outputs the waveguide calibration kit open capacitance C3 term on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit open reference frequency on the indicated channel. Outputs the waveguide calibration kit open reference frequency on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit open loss value on the indicated channel. Outputs the waveguide calibration kit open loss value on the indicated channel.
Sets the waveguide calibration kit open offset on the indicated channel. Outputs the waveguide calibration kit open offset on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
The command sets the waveguide calibration kit OPEN S1P file settings ON/OFF on the selected port of the indicated channel. ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
The query outputs the waveguide calibration kit OPEN S1P file settings on the selected port of the indicated channel.
Sets the waveguide calibration kit serial number on the indicated channel. Outputs the waveguide calibration kit serial number on the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Sets the waveguide calibration kit Short1 reference frequency on the indicated channel. Outputs the waveguide calibration kit Short1 reference frequency on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit Short1 loss value on the indicated channel. Outputs the waveguide calibration kit Short1 loss value on the indicated channel.
Sets the waveguide calibration kit short1 offset on the indicated channel. Outputs the waveguide calibration kit short1 offset on the indicated channel.
Sets the waveguide calibration kit Short2 reference frequency on the indicated channel. Outputs the waveguide calibration kit Short2 reference frequency on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit Short2 loss value on the indicated channel. Outputs the waveguide calibration kit Short2 loss value on the indicated channel.
Sets the waveguide calibration kit short2 offset on the indicated channel. Outputs the waveguide calibration kit label short2 offset on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit Short3 reference frequency on the indicated channel. Outputs the waveguide calibration kit Short3 reference frequency on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit Short3 loss value on the indicated channel. Outputs the waveguide calibration kit Short3 loss value on the indicated channel.
Sets the waveguide calibration kit short3 offset length on the indicated channel. Outputs the waveguide calibration kit short3 offset length on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the waveguide calibration kit sliding load minimum frequency on the indicated channel. Outputs the waveguide calibration kit sliding load minimum frequency on the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.